ge healthcare llcmedicalHigh PriorityActive

SIGNA Premier systems

SIGNA Premier systems was recalled because whole-body mr scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.. If you have this product, stop using it until you verify your lot or model details.

By RecallRadar Editorial Team

Sourced from FDA. Last verified April 15, 2026.

On this page

Free checker

Check if you qualify

See whether your device may qualify for a free repair or replacement program before you spend money.

Check if you qualify

Key facts

Date announced
April 8, 2026
Source agency
FDA

This recall involves safety concerns. Take action promptly.

Plain-English Summary

SIGNA Premier systems was recalled because whole-body mr scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.. If you have this product, stop using it until you verify your lot or model details.

Is This Recall Still Active?

Status: Yes - Active

This recall is currently active. If your device matches the affected models, follow the official instructions as soon as possible.

What's Being Recalled

Quoted source excerpt from the official notice.

Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.

Quick Info

Key facts from the official notice, kept in the article flow for easy cross-checking.

Brand
Ge healthcare llc
Category
medical
Source agency
FDA
Announced
April 8, 2026
Country
US
Status
Active

Who's Affected

Specific models and identifiers listed in the source notice:

Model #1

UDI-DI/GTIN: 00840682135269, 00195278010797. System ID: 415353UMR2, 415723SHMR19, 413253MR3T, 908725MR, 609652MR3T, M48823, 7168453T, 513584MR5, 513636EASTMR, 513636T1MR, 605333MRPRM, 832522CRMR1, 703858PRE2, 804675VA3T, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382, M4216469, 353745MR01, GON5655515, KRMR56474971, KRMR56203791, KRMR55433661, KRMR56344161, KRMR55774051, KRMR55914661, NOMRI26833321, 15974045, GON1882380, ESMRI27056983, ESMRI26960129, A203MR06, 690120MR01, 11524261, VNMRI26792525, VNMRI26592248, VNMRI26809677

Technical model codes are manufacturer identifiers. Match both the product name and model code from your device settings, label, or packaging.

What You Should Do

Follow the remedy path from the official notice using the exact model and serial details above.

  1. 1Stop using the affected product until you verify whether your item is impacted.
  2. 2Check the official recall notice and match your model, lot, or serial details.
  3. 3Follow the official instructions for repair, replacement, refund, or disposal.
Open Official Recall Notice

Source & Verification

This page summarizes the official notice and keeps key facts easy to verify.

Official source
FDA
Announced by source
April 8, 2026
Last reviewed by RecallRadar
April 15, 2026

Frequently Asked Questions

Am I affected by this recall?

Check the affected models section on this page. The official notice currently lists UDI-DI/GTIN: 00840682135269, 00195278010797. System ID: 415353UMR2, 415723SHMR19, 413253MR3T, 908725MR, 609652MR3T, M48823, 7168453T, 513584MR5, 513636EASTMR, 513636T1MR, 605333MRPRM, 832522CRMR1, 703858PRE2, 804675VA3T, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382, M4216469, 353745MR01, GON5655515, KRMR56474971, KRMR56203791, KRMR55433661, KRMR56344161, KRMR55774051, KRMR55914661, NOMRI26833321, 15974045, GON1882380, ESMRI27056983, ESMRI26960129, A203MR06, 690120MR01, 11524261, VNMRI26792525, VNMRI26592248, VNMRI26809677.

What is the remedy for this recall?

Follow the official instructions for repair, replacement, refund, or disposal on the ge healthcare llc recall page.

Where can I verify recall details?

Use the official source link in this page to confirm dates, eligibility, and next steps: https://api.fda.gov/device/enforcement.json?search=recall_number:Z-1767-2026&limit=1

Keep browsing with these internal hubs: ge healthcare llc recall hub and Medical recall hub.

Official Resources

Use these primary sources for final eligibility and claim instructions.

View official notice from FDA

See an inaccuracy?

Help us improve the accuracy of this recall information.

500 characters remaining