Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat, Small, Cat. No. ARS1041101 2. TORNIER HRS Max Tuberosity Body No Coat, Small, Cat. No. ARS1041102 3. TORNIER HRS Max Trial, Cat. No. ARS10252
Affected products were incorrectly labeled as Tornier HRS Max products. These parts are compatible with the Tornier HRS system only, and are incompatible with the Tornier HRS Max system.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CSTM MANIFOLD KIT - TITUS, Medline SKU VASC1061; 2. CSTM MANIFOLD KIT - FRYE, Medline SKU VASC1074; 3. CRYO EP MANIFOLD KIT, Medline SKU VASC1198; 4. MANIFOLD KIT - 500PSI OFF, Medline SKU VASCSHPOFF1; 5. MANIFOLD KIT LOW PRESSURE ON, Medline SKU VASCSLPON1.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →COOK MEDICAL Wayne Pneumothorax Tray: Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
Apr 15, 2026
View details →MicroScan Neg MIC 3J REF C54814
Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics
Apr 15, 2026
View details →COOK MEDICAL Multipurpose Drainage Catheter: Reference Part Number ULT8.5-38-25-P-CLMS-8-RH-FS, Order Number G29775
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →COOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →NAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG,RA,-,PKG, Medline SKU 70085107; 4. SYRINGE,12ML,TR/FR,RA,W/RES, Medline SKU 70087007; 5. MTO,SYRINGE,12ML,TR/FG,RA,W/RES, Medline SKU 70087107; 6. SYRINGE,12ML,TR/FR,RA,W/O,RES,-,PKG, Medline SKU 70088007; 7. SYRINGE,10ML,PP/FR,RA,-,PKG, Medline SKU 70095007; 8. SYRINGE,10ML,PP/FG,RA,-,PKG, Medline SKU 70095107; 9. SYRINGE,12ML,W/RES,PP/FR,RA,PG, Medline SKU 70097007; 10. SYRINGE,12ML,PP/FG,RA,W/RES,-,PKG, Medline SKU 70097107; 11. SYRINGE,12ML,PP/FG,RA,W/O,RES, Medline SKU 70098107; 12. SYRINGE,20ML,TR/FR,RA,OEM,PG, Medline SKU 80075027; 13. SYRINGE,10ML,ANGIO,W/RA,-,OEM, Medline SKU 80085007; 14. SYRINGE,12ML,TR/FR,RA,OEM, Medline SKU 80087007; 15. SYRINGE,10ML,PP/FR,RA,-,OEM,-,PG, Medline SKU 80095007; 16. SYRINGE,10ML,PP/FG,RA,OEM,PG, Medline SKU 80095107; 17. SYRINGE,12ML,PP/FG,RA,-,OEM,-,PG, Medline SKU 80097107
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →COOK MEDICAL Spectrum¿ Central Venous Catheter Set: Reference Part Number C-UDLM-401J-ABRM-HC, Order Number G43983; Reference Part Number C-UDLM-401J-ABRM-HC-CAH, Order Number G24754; Reference Part Number C-UDLM-401J-LSC-ABRM-HC, Order Number G43986; Reference Part Number C-UDLM-401J-PED-ABRM-HC, Order Number G43984; Reference Part Number C-UDLM-401J-PED-ABRM-HC-CAH, Order Number G24752; Reference Part Number C-UDLM-401J-RSC-ABRM-HC, Order Number G43985; Reference Part Number C-UDLM-501J-ABRM-HC, Order Number G43995; Reference Part Number C-UDLM-501J-PED-ABRM-HC, Order Number G43996; Reference Part Number C-UDLM-501J-RSC-ABRM-HC, Order Number G43997; Reference Part Number C-UDLM-801J-ABRM-HC, Order Number G50817; Reference Part Number C-UQLM-1001J-ABRM-HC-RD, Order Number G51537; Reference Part Number C-UQLM-1001J-RSC-ABRM-HC-RD, Order Number G51539; Reference Part Number C-UTLM-501J-ABRM-HC, Order Number G44005; Reference Part Number C-UTLM-501J-ABRM-HC-CAH, Order Number G24803; Reference Part Number C-UTLM-501J-LSC-ABRM-HC, Order Number G44008; Reference Part Number C-UTLM-501J-PED-ABRM-HC, Order Number G44006; Reference Part Number C-UTLM-501J-RSC-ABRM-HC, Order Number G44007; Reference Part Number C-UTLM-501J-RSC-ABRM-HC-CAH, Order Number G24765; Reference Part Number C-UTLM-701J-ABRM-HC-RD, Order Number G47825; Reference Part Number C-UTLM-701J-ABRM-HC-RD-CAH, Order Number G24809; Reference Part Number C-UTLM-701J-ABRM-HC-RD-P, Order Number G53599; Reference Part Number C-UTLM-701J-LSC-ABRM-HC-RD, Order Number G47827; Reference Part Number C-UTLM-701J-RSC-ABRM-HC-RD, Order Number G47829; Reference Part Number C-UTLM-701J-RSC-ABRM-HC-RD-CAH, Order Number G24810.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: MINOR PLASTIC-LF, Medline Kit SKU # DYNJ86596
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →COOK MEDICAL Ciaglia Blue Rhino Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-FLEX-HC, Order Number G32092; Reference Part Number C-PTIS-100-HC, Order Number G12115.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: TAVR PACK, Medline SKU # DYNJ84472
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. MTS RH LFT HRT KIT, Medline SKU # 60010508; 2. MTS 3V W/GUIDE, Medline SKU # 60100701; 3. ANGIOGRAPHY KIT, Medline SKU # 60131209; 4. MTS LHK, Medline SKU # 60160075; 5. MTS LHK, Medline SKU # 60161954; 6. MTO LHK, Medline SKU # 60230148; 7. GENERAL SURGERY TRAY, Medline SKU # 00-500510I; 8. BIOBURDEN TEST KIT - 686, Medline SKU # BIOB686; 9. RADIO ANGIO UNIV HOSP PACK-LF, Medline SKU # DYNJ0376145M; 10. SPECIALS PACK-LF, Medline SKU # DYNJ0392706K; 11. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426O; 12. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426P; 13. CARDIAC CATH PACK, Medline SKU # DYNJ07953N; 14. CATH PACK-LF, Medline SKU # DYNJ0803367J; 15. E.P. TRAY, Medline SKU # DYNJ19343J; 16. CARDIAC CATH PACK, Medline SKU # DYNJ20898T; 17. CATH LAB TRAY, Medline SKU # DYNJ24250G; 18. CATH LAB PACK-LF, Medline SKU # DYNJ38333D; 19. CATH LAB PACK-LF, Medline SKU # DYNJ38333F; 20. EP IMPLANT PACK-LF, Medline SKU # DYNJ38910D; 21. ADULT CATH LAB PACK, Medline SKU # DYNJ40380J; 22. CRANIOTOMY PACK-LF, Medline SKU # DYNJ43403D; 23. LAP CHOLE PACK, Medline SKU # DYNJ48487F; 24. CATH LAB ANGIOGRAPHY PACK, Medline SKU # DYNJ50779G; 25. CATH LAB ANGIOGRAPHY PACK, Medline SKU # DYNJ50779I; 26. HYBRID, Medline SKU # DYNJ57049N; 27. SPECIALS PK, Medline SKU # DYNJ57713C; 28. SPECIALS PK, Medline SKU # DYNJ57713D; 29. SPECIALS PK, Medline SKU # DYNJ57713F; 30. CARDIAC CATH LAB, Medline SKU # DYNJ57748A; 31. TAVR PACK, Medline SKU # DYNJ58771G; 32. CATH PACK GWV, Medline SKU # DYNJ59229F; 33. HEART CATH PACK, Medline SKU # DYNJ61294F; 34. CATH LAB PACK, Medline SKU # DYNJ61353C; 35. DEVICE IMPLANT PACK, Medline SKU # DYNJ64636A; 36. CUSTOM PROCEDURE KIT, Medline SKU # DYNJ65277A; 37. GENERAL VASCULAR PACK RFID-V2, Medline SKU # DYNJ66442F; 38. CARDIAC CATH PACK, Medline SKU # DYNJ69845C; 39. OMF MINOR PACK, Medline SKU # DYNJ83991C; 40. HEART CATH PACK, Medline SKU # DYNJ84133B; 41. SPECIAL PROCEDURES, Medline SKU # DYNJ85425B; 42. PRIMARY PACK, Medline SKU # DYNJT3093; 43. CARDIAC CATH PACK, Medline SKU # DYNJT3522A; 44. CARDIAC CATH PACK, Medline SKU # DYNJT3522B; 45. TOTAL KNEE PACK, Medline SKU # DYNJT4069; 46. CARDIAC CATH PACK, Medline SKU # DYNJT5685; 47. CATH PACK, Medline SKU # DYNJT6544; 48. PACK CARDIAC CATH, Medline SKU # DYNJT7591; 49. OBL CATH VEIN PACK, Medline SKU # DYNJT7863; 50. TUMESCENT SYRINGE KIT, Medline SKU # DYNJTUMSYR; 51. CSTM LH KIT, Medline SKU # VASC1054; 52. LHK, Medline SKU # VASC1130A; 53. MANIFOLD KIT, Medline SKU # VASC1137A; 54. LHK, Medline SKU # VASC1299A; 55. LHK, Medline SKU # VASC1399B; 56. LHK, Medline SKU # VASC1403; 57. IR 2 PORT, Medline SKU # VASC1404; 58. LHK, Medline SKU # VASC1405; 59. LHK, Medline SKU # VASC1455A; 60. TWO PORT MANIFOLD KIT, Medline SKU # VASC1496; 61. RHK, US, NAMIC MANIFOLD, Medline SKU # VASC1502; 62. LHK, Medline SKU # VASC1555; 63. 4 VALVE,PERC,MANIFOLD,KIT, Medline SKU # VASC1609; 64. LHK, Medline SKU # VASC1668; 65. LHK, Medline SKU # VASC1694; 66. LHK, Medline SKU # VASC1805; 67. TVR, Medline SKU # VASC1816.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. DELIVERY PACK, Medline SKU # DYNJ42892F; 2. DELIVERY PACK, Medline SKU # DYNJ42892G.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →See Luer Cap Set, MPC-130, set, administration, intravascular
Molded Products, Inc. is voluntarily conducting a medical device recall of the MPC-130 See Luer Cap Set based on a confirmed complaint of the threaded sleeve not being engaged and becoming un-attached.
Apr 15, 2026
View details →COOK MEDICAL Cook¿ Spectrum¿ Central Venous Catheter Tray: Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0003, Order Number G56424; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0018, Order Number G21053; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0042, Order Number G34914; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0047, Order Number G44431; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0048, Order Number G44432.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: GLOVE PACK 7.0 (15PK), Medline SKU # DYNJ63673D
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →COOK MEDICAL Cook Staged Extubation Set: Reference Part Number C-CAE-14.0-83-EXTU, Order Number G24073
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ANGIO PACK, Medline SKU # DYNJ57032D; 2. NEURO INTERVENTIONAL RADIOLOGY, Medline SKU # DYNJ64026C; 3. NEURO INERVENTIONAL RADIOLOGY, Medline SKU # VASC1650.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. SHOULDER ARTHROSCOPY PACK, Medline SKU # DYNJ30880M; 2. EXTREMITY PACK, Medline SKU # DYNJ64295A.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. HEAD AND NECK CDS-LF, Medline SKU # CDS780138K; 2. HEART CATH PACK-LF, Medline SKU # DYNJ36478B; 3. HEART CATH PACK-LF, Medline SKU # DYNJ36478B; 4. HEART CATH PACK-LF, Medline SKU # DYNJ36478B; 5. THYROIDECTOMY, Medline SKU # DYNJ910458; 6. LH 3 PORT W/O WASTEBAG, Medline SKU # VASC1081C; 7. LHK, Medline SKU # VASC1510; 8. LHK, Medline SKU # VASC1510.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →XTANT MEDICAL Xpress Pedicle Screw System Dual Lead 6.5 x 50 mm, Part Number X073-6555-DL.
Units from one lot were mislabeled with an incorrect part number, identifying the device as a shorter screw length than it physically is.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. NON VASCULAR PACK, Medline SKU # DYNJ68037C; 2. VEIN, Medline SKU # DYNJ89845; 3. VEIN, Medline SKU # DYNJ89845A.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →COOK MEDICAL Blue Rhino¿ G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703; Reference Part Number C-PTISY-100-UNL-HC-G-NA, Order Number G57706; Reference Part Number C-PTISY-100-UNS-HC-G-NA, Order Number G57704.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →COOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193; Reference Part Number C-TCCSB-500-UNI-01, Order Number G47665.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →NAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG,RA,-,PKG, Medline SKU 70085107; 4. SYRINGE,12ML,TR/FR,RA,W/RES, Medline SKU 70087007; 5. MTO,SYRINGE,12ML,TR/FG,RA,W/RES, Medline SKU 70087107; 6. SYRINGE,12ML,TR/FR,RA,W/O,RES,-,PKG, Medline SKU 70088007; 7. SYRINGE,10ML,PP/FR,RA,-,PKG, Medline SKU 70095007; 8. SYRINGE,10ML,PP/FG,RA,-,PKG, Medline SKU 70095107; 9. SYRINGE,12ML,W/RES,PP/FR,RA,PG, Medline SKU 70097007; 10. SYRINGE,12ML,PP/FG,RA,W/RES,-,PKG, Medline SKU 70097107; 11. SYRINGE,12ML,PP/FG,RA,W/O,RES, Medline SKU 70098107; 12. SYRINGE,20ML,TR/FR,RA,OEM,PG, Medline SKU 80075027; 13. SYRINGE,10ML,ANGIO,W/RA,-,OEM, Medline SKU 80085007; 14. SYRINGE,12ML,TR/FR,RA,OEM, Medline SKU 80087007; 15. SYRINGE,10ML,PP/FR,RA,-,OEM,-,PG, Medline SKU 80095007; 16. SYRINGE,10ML,PP/FG,RA,OEM,PG, Medline SKU 80095107; 17. SYRINGE,12ML,PP/FG,RA,-,OEM,-,PG, Medline SKU 80097107.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. LAPAROSCOPIC HERNIORRHAPHY DSC, Medline SKU # DYNJ63021B; 2. LAP PACK, Medline SKU # DYNJ64297B; 3. MINOR PROCEDURE PACK, Medline SKU # DYNJ82411.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007; 2. SYRINGE,8ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70084007; 3. MTS,SYRINGE,8ML,TR/FR,RA,NO,RES,-,P/G, Medline SKU 70084017; 4. SYRINGE,7,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80083007; 5. SYRINGE,8,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80084007; 6. MTO,SYRINGE,8,ML,TR/FR,RA,NO,RES,OEM,PG, Medline SKU 80084017.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →COOK MEDICAL Cook¿ Strange Bile Duct Stone Exploration Set: Reference Part Number C-CSRC-7.5D-40-STRANGE, Order Number G09108
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →COOK MEDICAL Approach CTO-18 Micro Wire Guide, Reference Part Number CMW-14-190-18G, Order Number G50788; Reference Part Number CMW-14-300-18G, Order Number G50792
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →COOK MEDICAL NCompass¿ Nitinol Stone Extractor: Reference Part Number C-NTSE-2.4-115-NC3, Order Number G36250; Reference Part Number C-NTSE-2.4-115-NCT4, Order Number G36251.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CATH LAB PACK-LF, Medline SKU 00-HCT055P; 2. INTERVENTIONAL CATH LAB PK-LF, Medline SKU DYNJ0376230Q; 3. ANGIOGRAPHY CATH LAB, Medline SKU DYNJ23532C; 4. ANGIO LEFT HEART PACK, Medline SKU DYNJ40632C; 5. ANGIO DRAPE PACK, Medline SKU DYNJ47952B; 6. ANGIOGRAPHIC PACK, Medline SKU DYNJ57516; 7. CATH LAB PACK, Medline SKU DYNJ58243B; 8. CATH LAB PACK, Medline SKU DYNJ65185; 9. CSU CATH MANIFOLD KIT, Medline SKU VASC1170A; 10. MANIFOLD KIT NO TRANSDUCER, Medline SKU VASC1184A.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →COOK MEDICAL Wayne Pneumothorax Set: Reference Part Number C-UTPT-1020-WAYNE-IMH, Order Number G56532; Reference Part Number C-UTPT-1400-WAYNE-112497-IMH, Order Number G56535.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N. in vitro diagnostic test for Placental Growth Factor
Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve. If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.
Apr 15, 2026
View details →COOK MEDICAL Approach CTO-6 Micro Wire Guide, Reference Part Number CMW-14-190-6G, Order Number G50786
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: SPECIAL PROCEDURE TRAY, Medline SKU # DYNJ80513B
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ROBOTIC PACK, Medline SKU # DYNJ58475F; 2. URO/GEN ROBOTIC PACK, Medline SKU # DYNJ60803J.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →COOK MEDICAL Spectrum¿ Central Venous Tray: Reference Part Number C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, Order Number G44800; Reference Part Number C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, Order Number G44809; Reference Part Number C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, Order Number G44806; Reference Part Number C-UTLMYJ-701J-RSC-ABRM-HC-IHI-FST-A-RD, Order Number G44807.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →COOK MEDICAL Thal-Quick Chest Tube Tray: Reference Part Number C-TQTSY-3200, Order Number G07242
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →COOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-HC-G-EU, Order Number G57685; Reference Part Number C-PTIS-100-HC-G-NA, Order Number G57682; Reference Part Number C-PTIS-100-UNL-HC-G-EU, Order Number G57687; Reference Part Number C-PTIS-100-UNS-HC-G-EU, Order Number G57686.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026
View details →Stryker Sustainability Solution Color Cuff 24" (Yellow) Quick Connect Dual Port Single Bladder REF 5921-024-235 Stryker Sustainability Solution Color Cuff 18" (Red), Quick Connect Dual Port Single Bladder REF 5921-018-135 Stryker Sustainability Solution Color Cuff 18" (Red) Dual Port Single Bladder REF 5921-018-235.
Due to potential for tourniquet cuffs to detach from the bladder at the welding connection and result in the cuff not being able to maintain pressure.
Apr 15, 2026
View details →NAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG,RA,-,PKG, Medline SKU 70085107; 4. SYRINGE,12ML,TR/FR,RA,W/RES, Medline SKU 70087007; 5. MTO,SYRINGE,12ML,TR/FG,RA,W/RES, Medline SKU 70087107; 6. SYRINGE,12ML,TR/FR,RA,W/O,RES,-,PKG, Medline SKU 70088007; 7. SYRINGE,10ML,PP/FR,RA,-,PKG, Medline SKU 70095007; 8. SYRINGE,10ML,PP/FG,RA,-,PKG, Medline SKU 70095107; 9. SYRINGE,12ML,W/RES,PP/FR,RA,PG, Medline SKU 70097007; 10. SYRINGE,12ML,PP/FG,RA,W/RES,-,PKG, Medline SKU 70097107; 11. SYRINGE,12ML,PP/FG,RA,W/O,RES, Medline SKU 70098107; 12. SYRINGE,20ML,TR/FR,RA,OEM,PG, Medline SKU 80075027; 13. SYRINGE,10ML,ANGIO,W/RA,-,OEM, Medline SKU 80085007; 14. SYRINGE,12ML,TR/FR,RA,OEM, Medline SKU 80087007; 15. SYRINGE,10ML,PP/FR,RA,-,OEM,-,PG, Medline SKU 80095007; 16. SYRINGE,10ML,PP/FG,RA,OEM,PG, Medline SKU 80095107; 17. SYRINGE,12ML,PP/FG,RA,-,OEM,-,PG, Medline SKU 80097107
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
View details →Artoura Breast Tissue Expanders Reference Numbers: TEXP110RH TEXP120RH TEXP120RUH TEXP130RH TEXP135RH TEXP140RH TEXP150RH Textured, High & Ultra High Profile, Integral Injection Dome, 300cc, 375cc, 535cc, 475cc, 500cc, 600cc, 750cc,
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
Apr 8, 2026
View details →ALLURA Xper FD10; System Code: (1) 722003, (2)722010, (3)722026;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Azurion 7 B12; System Code: (1)722067, (2)722225, (3)722235;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Cardinal Health Monoject Tuberculin Syringe, 1 mL Luer Lock, Catalog Number 1180100777
The outer packaging and blister pack labeling identify the product as 1mL Luer Lock Tuberculin Syringe (Product Code 1180100777), however the syringes contained within the packages are U-100 Insulin syringes.
Apr 8, 2026
View details →2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.
Apr 8, 2026
View details →Allura Xper FD10C; System Code: 722001;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation
Due to firm distributing a nonconforming product that supposed to be scrapped and not for human use.
Apr 8, 2026
View details →MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →Azurion 7 M20; System Code: (1)722079, (2)722224, (3)722234, (4)722282 (OUS only);
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →MEDLINE ANTERIOR HIP PACK DYNJ64672B
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Apr 8, 2026
View details →Vue Motion V12. Product Number: 1017979.
Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.
Apr 8, 2026
View details →Burlington Medical, BAT (Breast, Axilla and Thyroid) Coverage.
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →ALLURA Xper FD20; System Code: (1) 722006, (2)722012, (3)722028;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Burlington Medical, Kilt
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →Brand Name: da Vinci S, Si Tenaculum Forceps Instrument Product Name: da Vinci S, Si Tenaculum Forceps Instrument Model/Catalog Number: 420207 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: N/A
Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments
Apr 8, 2026
View details →Allura Xper FD20/15; System Code: 722058;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
Apr 8, 2026
View details →Stryker CHROMOPHARE Softlit Ring Surgical Light System REF: CH00000001
Surgical light assembly may not adequate support the weight of the ceiling cover.
Apr 8, 2026
View details →Allura Xper FD20/20; System Code: 722038;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Burlington Medical, Caps.
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →ALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Medline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT, DYKE1915
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Apr 8, 2026
View details →Burlington Medical, Thyroid Shield.
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →ALLURA Xper FD10/10; System Code: (1) 722005, (2)722011, (3)722027;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →MiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →eTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-149
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
Apr 8, 2026
View details →Brand Name: da Vinci S, Si Double Fenestrated Grasper Instrument Product Name: da Vinci S, Si Double Fenestrated Grasper Instrument Model/Catalog Number: 420189 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: NA
Due to increased complaints for broken/frayed grip cables for reusable instruments.
Apr 8, 2026
View details →Brand Name: Finesse BTK Multicath Product Name: Injectable Angioplasty Balloon Catheter Model/Catalog Number: FS252251502 Product Description: 5 French sheath, 0.014" guidewire compatible over-the-wire injectable angioplasty balloon catheter, balloon diameter 2.5mm, balloon length 225mm, catheter working length 150cm Component: NA
Potential for the balloon in the device to not meet burst specifications.
Apr 8, 2026
View details →Sterile Radiology Procedure Kits, Model Number DYNDH1491B
There is a potential breach in the sterile pouch packaging for one lot of radiology kits, which may compromise product sterility.
Apr 8, 2026
View details →Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Apr 8, 2026
View details →Allura Xper FD20/20 OR Table; System Code: 722039;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
Apr 8, 2026
View details →Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
Apr 8, 2026
View details →Blue Ventilator Adapter Module, single-patient use circuit, Product Code M07937. Accessory to Baxter Volara Respiratory Therapy system.
Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue injury from overexpansion (barotrauma) when the Volara device is used in the home care environment. This correction is providing the home caregiver specific instructions for use.
Apr 8, 2026
View details →Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: N/A
Due to increased complaints for broken/frayed grip cables for reusable instruments.
Apr 8, 2026
View details →eTRAX Needle Sensor - 16G(for Aurora Trackers), Part Number 667-158
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
Apr 8, 2026
View details →Burlington Medical, Blockers.
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
Apr 8, 2026
View details →Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Artoura Breast Tissue Expanders Reference Numbers: SDC-100UH SDC-110UH SDC-120UH SDC-130UH SDC-135UH SDC-140UH Smooth Ultra High Profile Tissue Expander, 350cc, 455cc, 535cc, 650cc, 700cc, 850cc
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
Apr 8, 2026
View details →Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00. Continuous home-use ventilator device.
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
Apr 8, 2026
View details →RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00
Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
Apr 8, 2026
View details →CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-107MH, SCPX-117MH,SCPX-127MH, SCPX-135MH, SCPX-146MH, SCPX-156MH Smooth Medium Height Tissue Expander, 275cc, 350cc, 450cc, 550cc, 650cc, 800cc
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
Apr 8, 2026
View details →Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →MiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →Burlington Medical, Frontal Aprons
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →MiniMed 530G Insulin Pump (MMT-551, MMT-751)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146B 7. LEX THORACIC, DYNJ902016K DYNJ902016L 8. ORGAN PROCUREMENT, DYNJ908686 9. PULMONARY PACK, DYKE1859 10. THORACIC, DYNJ901666L 11. THORACIC ROBOTICS, DYNJ908777A 12. THORACOTOMY/THORACOSCOPY, DYNJ900482N DYNJ900482O 13. XPS, DYNJ907605
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Apr 8, 2026
View details →Azurion 3 M15; System Code: (1)722064, (2)722222, (3)722230, (4)722280 (OUS Only);
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Jiffy Original Composite Polisher Cups (Coarse) 12pk, REF 7011.
Composite polisher cups may crumble and break apart easily, which could cause a delay in patient treatment.
Apr 8, 2026
View details →eTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-156
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
Apr 8, 2026
View details →ALLURA Xper FD20 OR Table; System Code: (1) 722015, (2)722023, (3)722035;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Burlington Medical, Protective Sleeves.
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation
Due to firm distributing a nonconforming product that supposed to be scrapped and not for human use.
Apr 8, 2026
View details →ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC
Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
Apr 8, 2026
View details →Artoura Breast Tissue Expanders Reference Numbers: SDC-100H SDC-110H SDC-120H SDC-130H SDC-135H SDC-140H SDC-150H SDC-155H Smooth High Profile Tissue Expander, 225cc, 300cc, 375cc, 475cc, 500cc, 600cc, 750cc, 850cc
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
Apr 8, 2026
View details →Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146A 4. RIGID KIT, DYNDA2138A
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Apr 8, 2026
View details →Brand Name: da Vinci S, Si Grasping Retractor Instrument Product Name: da Vinci S, Si Grasping Retractor Instrument Model/Catalog Number: 420278 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: NA
Due to increased complaints for broken/frayed grip cables for reusable instruments.
Apr 8, 2026
View details →Allura Xper FD20/10; System Code: 722029;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Purge Cassette, sold within Impella Pump Sets and Individually Packaged. Labeled and distributed in the below configurations and corresponding product codes. 1. Purge Cassette, 5 Pack; Product Code: 0043-0003. 2. Impella RP US Pump Set; Product Code: 004334. 3. Purge Cassette, Packaged; Product Code: 0043-0002. Also included inside Product Code 0043-0003 above. Product Code 0043-0002 also includes innermost packaged product with Product Code 0043-0001. 4. Impella 2.5 Set; Product Code: 004413 (Customer facing code 005040). 5. Impella RP Pump Set, EU; Product Code: 0046-0011. 6. Impella CP Pump Set, EU; Product Code: 0048-0002. 7. Impella CP Pump Set BR; Product Code: 0048-0002-BR. 8. Impella CP Smart Assist Set, EU; Product Code: 0048-0014 (Also inclusive of codes 0048-0047 and 1000402 listed below). 9. Impella CP Smart Assist Set, Canada; Product Code: 0048-0044. 10. Impella CP Smart Assist Set APAC; Product Code: 0048-0047 (inclusive under code 0048-0014). 11. Impella 5.0 IMC Pump Set EU; Product Code: 005060. 12. Impella 5.5 with SmartAssist Set, EU; Product Code: 0550-0002. 13. Impella 5.5 with SmartAssist Set, CA; Product Code: 0550-0004. 14. 5.5 Accessories AU; Product Code: 1000362. 15. Impella CP Smart Assist Set; Product Code: 1000402 (inclusive under code 0048-0014).
Increased risk of purge leaks with Generation 1 purge cassettes.
Apr 8, 2026
View details →Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-157LH Smooth Low Height Tissue Expander, 650cc
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
Apr 8, 2026
View details →Burlington Medical, Kilt Blockers.
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: 354-9113, 3549313, 354-9314 SILTEX Low Height Tissue Expander Style (9100, 450cc), (9300, 450cc), (9300, 550cc)
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
Apr 8, 2026
View details →Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT
Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
Apr 8, 2026
View details →Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.
Apr 8, 2026
View details →ALLURA Xper FD10F; System Code: 722002;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Burlington Medical, Demi Half Aprons.
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →Automated Impella Controller (AIC), labeled as the
Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.
Apr 8, 2026
View details →Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
Apr 8, 2026
View details →MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →eTRAX Needle System Starter Kit 14G (for Aurora Trackers), Part Number 667-150
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
Apr 8, 2026
View details →ALLURA Xper FD20 Biplane OR Table; System Code: (1) 722020, (2)722025;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Burlington Medical, Vest
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem
Oxoid Ltd., part of Thermo Fisher Scientific, confirmed that Microbact 12L Kit 20 Tests MB1128A lot 4494873 may identify incorrect organism due to inconclusive result from ambiguous color reactions in wells.
Apr 8, 2026
View details →Burlington Medical, Half Aprons.
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →Azurion 5 M12; System Code: (1)722227, (2)722231;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Paradigm Insulin Pump (MMT-712, MMT-715)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
Apr 8, 2026
View details →Airway Exam Kit, DYKE1796
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Apr 8, 2026
View details →Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →eTRAX Needle Sensor - 14G(for Aurora Trackers), Part Number 667-157
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
Apr 8, 2026
View details →Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →BD Kiestra" ReadA; Catalog No.: 446948.
In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.
Apr 8, 2026
View details →CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-113TH, SCPX-123TH, SCPX-140TH, SCPX-146TH Smooth Tall Height Tissue Expander, 350cc, 450cc, 650cc, 750cc
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
Apr 8, 2026
View details →MiniMed 620G Insulin Pump (MMT-1750)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →Allura Xper FD20/15 OR Table; System Code: 722059;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Brand Name: da Vinci S, Si Mega Needle Driver Instrument Product Name: da Vinci S, Si Mega Needle Driver Instrument Model/Catalog Number: 420194 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: NA
Due to increased complaints for broken/frayed grip cables for reusable instruments.
Apr 8, 2026
View details →2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US.
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.
Apr 8, 2026
View details →eTRAX Needle Sensor - 18G (for Aurora Trackers), Part Number 667-159
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
Apr 8, 2026
View details →Burlington Medical, Leg Wraps.
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →ALLURA Xper FD20 Biplane; System Code: (1) 722008, (2)722013;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Apr 8, 2026
View details →Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →MiniMed 670G Insulin Pump (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
Apr 8, 2026
View details →Burlington Medical, Wrap Aprons.
Potential for attenuation degradation over time, decreasing the lifespan.
Apr 8, 2026
View details →MiniMed 720G Insulin Pump (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Apr 8, 2026
View details →THORACIC ROBOTS, DYNJ908777B
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Apr 8, 2026
View details →Brand Name: da Vinci S, Si Permanent Cautery Hook Instrument Product Name: da Vinci S, Si Permanent Cautery Hook Instrument Model/Catalog Number: 420183 Software Version: NA Product Description: EndoWrist Instruments or da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: NA
Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments
Apr 8, 2026
View details →SIGNA Premier systems
Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
Apr 8, 2026
View details →I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21234-4, 21235-4. 2. I.T.S. UOL (Ulna Osteotomy Locking) System, Ulna Osteotomy Plate; Article Numbers: 21231-6. 3. UOL (Ulna Osteotomy Locking) System, Cancellous Screw, Locking; Article Numbers: 37302-18, 37302-20.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Philips L17-5 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Apr 1, 2026
View details →Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
Apr 1, 2026
View details →HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31257-30, 31307-18, 31307-32, 31307-34, 31307-36, 31307-38, 31358-30, 31437-30, 31437-38, 31437-40, 31437-50, 31757-100, 31757-120, 31757-45, 31757-80, 31757-85, 31757-90.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee System
Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
Apr 1, 2026
View details →Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System
Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
Apr 1, 2026
View details →BioFlo DuraMax Catheter, REF: H965103028011/A, H965103028021/A, H965103028021/EU, H965103028031/A, H965103028031/EU, H965103028041/A, H965103028041/EU, H965103028051/A, H965103028051/EU, H965103028061/A, H965103028061/EU, H965103028070/A, H965103028070/EU, H965103028080/A, H965103028080/EU, H965103028090/A, H965103028090/EU, H965103028161/A, H965103028181/A, H965103028191/A, H965103028191/EU, H965103028201/A, H965103028211/A, H965103028220/A, H965103028230/A, H965103028240/A, H965103038181/A
16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, and procedure delay.
Apr 1, 2026
View details →Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd
If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.
Apr 1, 2026
View details →Olympus Endoscope Air/Water Valve - MAJ-1444. Model Number: MAJ-1444.
The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
Apr 1, 2026
View details →Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;
A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.
Apr 1, 2026
View details →Brand Name: Olympus Everest Bipolar Cutting Forceps, 5 mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3005 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Apr 1, 2026
View details →I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 21602-5, 21602-6, 21602-7, 21602-8, 21603-3, 21603-4, 21603-5, 21603-6, 21603-7, 21603-8. 2. Fibula Plate 3.5mm; Article Numbers: 21225-4, 21225-6, 21225-8, 21225-10, 21225-12, 21226-4, 21226-6, 21226-8, 21226-10, 21226-12. 3. Universal Tubular Plate; Article Numbers: 21601-5, 21601-6, 21601-7, 21601-8, 21601-10, 21601-12, 21601-14, 21601-16.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Model/Catalog Number: DIB00 Software Version: N/A Product Description: The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. The lens is preloaded and pre-assembled in the delivery system. The TECNIS SIMPLICITY" Delivery System Model DIB00 contains the TECNISEyhance" IOL, which is a one-piece, foldable, posterior chamber lens with an overall diameter of 13.0 mm and an optic diameter of 6.0 mm. The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. Component: No
Due to identifying a limited quantity of intraocular lenses that may have a haptic that sticks to the optic which prevents the lens from unfolding as consistently as expected.
Apr 1, 2026
View details →DuraMax Chronic Hemodialysis Catheter, REF: H787103028015/A, H787103028025/A, H787103028031/A, H787103028035/A, H787103028041/A, H787103028045/A, H787103028051/A, H787103028055/A, H787103028061/A, H787103028075/A, H787103028085/A, H787103028095/A, H787103028165/A, H787103028181/A, H787103028185/A, H787103028191/A, H787103028195/A, H787103028205/A, H787103028215/A, H787103028225/A, H787103028235/A
16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, and procedure delay.
Apr 1, 2026
View details →Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-10. 2. Clavicle Plate; Article Numbers: 21121-6, 21121-8. 3. Clavicle Plate, 3.5mm; Article Numbers: 21122-10, 21122-6, 21122-8. 4. Clavicle Plate, Lateral; Article Numbers: 21123-4, 21124-4, 21127-6, 21128-6.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Brand Name: Humeral Head Plate with Angular Stability, with the
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Philips Mini Multi TEE Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Apr 1, 2026
View details →Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
Apr 1, 2026
View details →Philips L12-5 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Apr 1, 2026
View details →Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"
Elevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation, delayed wound healing, or systemic effects such as fever and hypotension. In severe cases, septic shock may result.
Apr 1, 2026
View details →Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm laparoscopic cannula. Coagulation is achieved using electrosurgical energy under laparoscopic visualization. The device is intended to be used with the Gyrus ACMI G400 Workstation only.
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Apr 1, 2026
View details →MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
Due to the likely presence of contamination in well(s).
Apr 1, 2026
View details →Philips C9-4 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Apr 1, 2026
View details →Volar Radius Plate with Angular Stability with the
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probe Model/Catalog Number: ES-100V3 (device); BT2M20S8C (2 MHz fetal probe). Software Version: N/A Product Description: Hand-held, battery-powered ultrasonic Doppler blood-flow detector with LCD numeric display, used with interchangeable probes to detect vascular blood-flow signals and display heart rate; affected units were packaged with a 2 MHz BT2M20S8C fetal probe and labeled for both vascular and fetal applications.
Product labeling includes a fetal indication for use that is not cleared under its 510(k).
Apr 1, 2026
View details →Flexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use
Probes may rupture/burst during activation
Apr 1, 2026
View details →Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Apr 1, 2026
View details →Philips X7-2 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Apr 1, 2026
View details →Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Apr 1, 2026
View details →Philips 3D6-2 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Apr 1, 2026
View details →Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 21094-8. 3. Pilon Plate, Small; Article Numbers: 21093-4, 21093-6.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →FR.O.H. Calcaneus Repair System with the
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.
Apr 1, 2026
View details →Philips OMNI III TEE Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Apr 1, 2026
View details →Embrace Drill Tower, Standard/Lateral (25mm); Item Number: 645-081/63;
The surgical technique for the Embrace Shoulder Instruments - Drill Tower has mixed-up article numbers in the overview of the instruments. However, the individual surgical steps described in the surgical technique are correct.
Apr 1, 2026
View details →Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient).
The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.
Apr 1, 2026
View details →MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61
Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or hyperglycemia due to inaccuracies in insulin delivery (either under-delivery, over-delivery, or cessation of delivery).
Apr 1, 2026
View details →I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article numbers. 1. DFL (Distal Femur Locking) Distal Femur Plate; Article Numbers: 21301-5, 21301-7, 21301-9, 21301-13, 21302-5, 21302-7, 21302-9, 21302-13. 2. PTL (Proximal Lateral Tibia Locking) Proximal Lateral Tibia Plate; Article Numbers: 21321-7, 21321-12, 21322-7, 21322-12. 3.PTL (Proximal Lateral Tibia Locking) Proximal Medial Tibia Plate; Article Numbers: 21325-4, 21325-7.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →16F Dual Valved Splittable Sheath Introducer (bulk, non-sterile), REF: FCL-174-00/B
16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, and procedure delay.
Apr 1, 2026
View details →Brand Name: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model/Catalog Number: PK-CF0533 Product Description: An assembly of devices intended to generate and apply radio-frequency (RF) alternating current to soft tissues for cutting and coagulation during an endoscopic or open surgical procedure.
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Apr 1, 2026
View details →Olympus Endoscope Reprocessor OER-Elite. Model Number: N5789340.
The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
Apr 1, 2026
View details →Embrace Drill Tower, Wedged (25mm); Item Number: 645-081/62;
The surgical technique for the Embrace Shoulder Instruments - Drill Tower has mixed-up article numbers in the overview of the instruments. However, the individual surgical steps described in the surgical technique are correct.
Apr 1, 2026
View details →Philips OMNI II TEE Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Apr 1, 2026
View details →Olympus Endoscope Reprocessor OER-Pro. Model Number: N3058140.
The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
Apr 1, 2026
View details →Philips X3-1 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Apr 1, 2026
View details →Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00. Continuous home-use ventilator device.
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
Apr 1, 2026
View details →BD Alaris" System with Guardrails" Suite MX with Point of Care Unit
Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units
Apr 1, 2026
View details →Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP
There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.
Apr 1, 2026
View details →Philips S4-1 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Apr 1, 2026
View details →Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Apr 1, 2026
View details →ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C
16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, and procedure delay.
Apr 1, 2026
View details →I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Plate; Article Numbers: 21105-6, 21105-8. 4. Straight Plate, 1.5mm; Article Numbers: 21104-4, 21104-5, 21104-6, 21104-7, 21104-8, 21104-9. 5. Straight Plate, 2.0mm; Article Numbers: 21101-5, 21101-6, 21101-7, 21101-8, 21101-9, 21101-11, 21101-13. 6. Straight Plate, 3.5mm; Article Number: 21102-6. 7. Straight Plate, 4.5mm; Article Numbers: 21103-8, 21103-11, 21103-10, 21103-11, 21103-12, 21103-13, 21103-8, 21103-9.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 37351-20-N, 37351-22-N, 37351-24-N, 37351-26-N, 37351-28-N, 37351-30-N, 37351-32-N, 37351-34-N, 37351-36-N, 37351-38-N, 37351-40-N, 37351-42-N, 37351-44-N, 37351-46-N, 37351-48-N, 37351-50-N, 37351-55-N, 37351-60-N, 37351-65-N, 37351-70-N, 37351-75-N, 37351-80-N, 37351-85-N. 2. Humeral Plate, Distal; Article Numbers: 21241-5, 21241-7, 21242-7, 21243-10, 21244-9.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.
The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
Apr 1, 2026
View details →Flexible Cryoprobe (OD 2.4mm, L1.15mm) REF: 20402-411. For surgical use
Probes may rupture/burst during activation
Apr 1, 2026
View details →HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1. Cortical Screw; Article Numbers: 32151-6, 32151-8, 32151-9, 32151-10, 32151-11, 32151-12, 32151-13, 32151-14, 32151-15, 32151-16, 32151-18, 32151-20. 2. HLS Plate, 2 T-Shape; Article Numbers: 21505-4, 21506-4. 3. HLS Plate, 3 T-Shape; Article Numbers: 21507-4, 21508-4. 4. HLS Plate, Square; Article Numbers: 21517. 5. HLS Plate, Straight; Article Numbers: 21502, 21503-4, 21503-6, 21503-8, 21503-10, 21504-6, 21504-8, 21504-10. 6. HLS Plate, Y-Shape; Article Numbers: 21513-3, 21513-5, 21514-3, 21514-5. 7. Locking Stabilization Screw; Article Numbers: 37151-5, 37151-8, 37151-9, 37151-10, 37151-12, 37151-13, 37151-15, 37151-16, 37151-18, 37151-24, 37251-8, 37251-9, 37251-11, 37251-12, 37251-13, 37251-14, 37251-16, 37251-20.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below descriptions and corresponding article numbers. 1. Anterior Column Plate; Article Numbers: 21218-16, 21219-16. 2. Anterior Column Plate; Article Numbers: 21218-14, 21219-14. 3. Biplanar 2-Column Plate; Article Numbers: 21216-10, 21216-12, 21217-10, 21217-12. 4. Cancellous Screw; Article Numbers: 30591-36, 30591-70, 30591-75, 30591-80, 30591-85, 30591-90. 5. Cancellous Screw, Locking; Article Numbers: 37422-12-N, 37422-14-N, 37422-16-N, 37422-18-N, 37422-20-N, 37422-22-N, 37422-24-N, 37422-26-N, 37422-28-N, 37422-30-N, 37422-32-N, 37422-34-N, 37422-36-N, 37422-38-N, 37422-40-N, 37422-42-N, 37422-44-N, 37422-46-N, 37422-48-N, 37422-50-N, 37422-55-N, 37422-60-N, 37422-65-N, 37422-70-N, 37422-75-N, 37422-80-N, 37422-85-N, 37422-90-N, 37422-95-N, 37422-100-N, 37422-105-N, 37422-110-N, 37422-115-N, 37422-120-N, 37592-18, 37592-20, 37592-22, 37592-24, 37592-26, 37592-28, 37592-30, 37592-32, 37592-34, 37592-36, 37592-38, 37592-40, 37592-42, 37592-46, 37592-48, 37592-50, 37592-54, 37592-56, 37592-58, 37592-60, 37592-65, 37592-70, 37592-75, 37592-80, 37592-85, 37592-90. 6. Cannulated Screw, Cancellous; Article Numbers: 31851-60, 31851-65, 31851-70, 31851-75, 31851-80, 31851-85, 31851-90, 31851-95, 31851-100, 31851-105, 31851-110, 31851-115, 31851-120, 31851-125, 31851-130, 31851-135, 31851-140, 31851-145, 31851-150, 31851-155, 31851-160, 31851-165, 31851-170, 31851-175, 31851-180, 31851-185, 31851-190, 31851-195, 31851-200, 31854-60, 31854-65, 31854-70, 31854-75, 31854-80, 31854-85, 31854-90, 31854-95, 31854-100, 31854-105, 31854-110, 31854-115, 31854-120, 31854-125, 31854-130, 31854-135, 31854-140, 31854-145, 31854-150, 31854-155, 31854-160, 31854-165, 31854-170, 31854-175, 31854-180, 31854-185, 31854-190, 31854-195, 31854-200. 7. Cortical Screw; Article Numbers: 32351-10, 32351-12, 32351-14, 32351-16, 32351-18, 32351-20, 32351-22, 32351-24, 32351-26, 32351-28, 32351-30, 32351-32, 32351-34, 32351-36, 32351-38, 32351-40, 32351-42, 32351-44, 32351-46, 32351-48, 32351-50, 32351-55, 32351-60, 32351-65, 32351-70, 32351-75, 32351-80, 32351-85, 32352-16, 32352-18, 32352-20, 32352-22, 32352-24, 32352-26, 32352-28, 32352-30, 32352-32, 32352-34, 32352-36, 32352-38, 32352-40, 32352-42, 32352-44, 32352-46, 32352-48, 32352-50, 32352-55, 32352-60, 32352-65, 32352-70, 32352-75, 32352-80, 32352-85, 32352-90, 32352-95, 32352-100, 32352-105, 32352-110, 32352-115, 32352-120, 32455-18, 32455-20, 32455-22, 32455-24, 32455-26, 32455-28, 32455-30, 32455-32, 32455-34, 32455-36, 32455-38, 32455-40, 32455-42, 32455-44, 32455-46, 32455-48, 32455-50, 32455-52, 32455-54, 32455-56, 32455-58, 32455-60, 32455-65, 32455-70, 32455-75, 32455-80, 32455-85, 32455-90, 32475-32. 8. Cortical Screw, Locking; Article Numbers: 37455-18, 37455-20, 37455-22, 37455-24, 37455-26, 37455-28, 37455-30, 37455-32, 37455-34, 37455-36, 37455-38, 37455-40, 37455-42, 37455-44, 37455-46, 37455-48, 37455-50, 37455-52, 37455-54, 37455-56, 37455-60, 37455-65, 37455-70, 37455-75, 37455-80, 37455-85, 37455-90. 9. Curved Plate; Article Numbers: 211911-4, 211911-5, 211911-6, 211911-7, 211911-8, 211911-10, 211911-12, 211911-14, 211911-16, 211911-20. 10. Posterior Column Plate, Left; Article Numbers: 212121-6. 11. Posterior Column Plate, Right; Article Numbers: 212131-6. 12. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-SM. 13. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-LA. 14. Posterior Wall Plate, Extended, Right; Article Numbers: 212111-LA, 212111-SM. 15. Posterior Wall Plate, Left; Article Numbers: 212141-7, 212141-8. 16. Posterior Wall Plate, Right; Article Numbers: 212151-7, 212151-8. 17. PRS Plate; Article Numbers: 21194-6, 21194-7, 21194-8, 21195-10, 21195-12, 21195-7. 18. Quadrilateral Buttress Plate; Article Numbers: 21208-13, 21209-13. 19. Quadrilateral Buttress Plate; Article Numbers: 21208-11, 21209-11. 20. Rim Plate, Left; Article
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →CentrosFLO Hemodialysis Catheters, REF: CENFP15K/A, CENFP17K/A, CENFP19K/A, CENFP23K/A, CENFP27K/A, CENFP31K/A, CENFT15K, CENFT15K/D, CENFT17K, CENFT17K/A, CENFT17K/D, CENFT19K, CENFT19K/A, CENFT19K/B, CENFT19K/D, CENFT23K, CENFT23K/A, CENFT23K/B, CENFT23K/D, CENFT27K, CENFT27K/A, CENFT27K/B, CENFT27K/D, CENFT31K, CENFT31K/A, CENFT31K/B, CENFT31K/D
16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, and procedure delay.
Apr 1, 2026
View details →Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Apr 1, 2026
View details →I.T.S. Screw System with the below descriptions and corresponding article numbers. 1. Cancellous Screw, Cannulated; Article Numbers: 31404-16, 31404-18, 31404-20, 31404-22, 31404-24, 31404-26, 31404-28, 31404-30, 31404-32, 31404-34, 31404-36, 31404-38, 31404-40, 31404-42, 31404-44, 31404-46, 31404-48, 31404-50, 31404-55, 31404-60, 31404-65, 31404-70, 31404-75, 31404-80, 31457-30, 31457-36, 31457-40, 31457-46, 31457-50, 31457-55, 31457-60, 31652-100, 31652-105, 31652-110, 31652-115, 31652-120, 31652-25, 31652-30, 31652-35, 31652-40, 31652-45, 31652-55, 31652-60, 31652-65, 31652-70, 31652-75, 31652-80, 31652-85, 31652-90, 31652-95, 31731-100, 31731-105, 31731-110, 31731-115, 31731-120, 31731-50, 31731-60, 31731-65, 31731-70, 31731-75, 31731-80, 31731-85, 31731-90, 31731-95, 31732-60, 31734-100, 31734-105, 31734-110, 31734-115, 31734-120, 31734-45, 31734-50, 31734-60, 31734-65, 31734-70, 31734-75, 31734-80, 31734-85, 31734-90, 31734-95. 2. Cortical Screw, Cannulated; Article Numbers: 31357-44, 31357-22, 31357-48, 31277-18, 31357-20, 31357-22, 31357-24, 31357-26, 31357-28, 31357-30, 31357-32, 31357-34, 31357-36, 31357-38, 31357-40, 31357-42, 31357-44, 31357-46, 31357-48, 31357-60, 31357-65, 31357-70. 3. Washer; Article Number: 36851. 4. Washer, Concave; Article Numbers: 36432, 36653, 36732. 5. Washer, Flat; Article Numbers: 36431, 36651, 36653, 36731.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Philips S5-2 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Apr 1, 2026
View details →FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Numbers: 37303-10, 37303-12, 37303-14, 37303-16, 37303-18, 37303-20, 37303-22, 37303-24, 37303-26, 37303-28, 37303-30. 2. FLS - Foot Locking Plates System, Cortical Screw; Article Numbers: 32271-10, 32271-12, 32271-14, 32271-16, 32271-18, 32271-20, 32271-22, 32271-24, 32271-26, 32271-28, 32271-32, 32271-34. 3. FLS - Foot Locking Plates System, Cortical Stabilization Screw; Article Numbers: 37304-10, 37304-12, 37304-14, 37304-16, 37304-18, 37304-20, 37304-22, 37304-24. 4. FLS - Foot Locking Plates System, FLS Plate, Curved; Article Numbers: 21413-6. 5. FLS - Foot Locking Plates System, FLS Plate, H-Shape; Article Numbers: 21431-ME. 6. FLS - Foot Locking Plates System, FLS Plate, L-Shape; Article Numbers: 21482, 21483. 7. FLS - Foot Locking Plates System, FLS Plate, Straight; Article Numbers: 21421-5. 8. FLS - Foot Locking Plates System, Stabilization Screw; Article Numbers: 37241-10, 37241-12, 37241-14, 37241-16, 37241-18, 37241-20, 37241-22, 37241-24, 37241-26, 37241-28, 37241-30. 9. HOL - Hallux Osteotomy Locking Plate, Hallux Osteotomy Plate; Article Numbers: 21015-08, 21015-09.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Apr 1, 2026
View details →Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
Apr 1, 2026
View details →Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILEEO. For surgical use
Probes may rupture/burst during activation
Apr 1, 2026
View details →Convatec, EsteemBody Drainable Pouch REF:423657 10-35mm
Drainable large pouch may leak due to manufacturing issue.
Mar 25, 2026
View details →Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package
Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
Mar 25, 2026
View details →Centricity Universal Viewer Software Versions 7.0 through 7.0 Sp2.0.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 5826659-027, 5826659-029, 5826659-032, 5826659-033, 5826659-028, 5826659-034, 5826659-030, 5826659-019, 5826659-022, 5826659-026, 5826659-023, 5826659-031, 5826659-025, 5826659-002, 5826659-021, 5826659-037, 5826659-014, 5826659-016, 5826659-013
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
Mar 25, 2026
View details →Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-026, 2088026-115, 2088026-043, K2042VJED, 2088026-132, 2088026-003, 2088026-110, 2089629-003, 2066908-136, 2066908-150
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
Mar 25, 2026
View details →Artegraft Vascular Graft; REF#: AG740;
Labeling mix-up resulting in the incorrect lot outer packaging of product.
Mar 25, 2026
View details →Brand Name: Aptima HPV Assay Product Name: Aptima HPV Assay Model/Catalog Number: 303585 Product Description: Aptima HPV assay Component: No
Due to product exhibiting potential to generate either invalid or false negative results.
Mar 25, 2026
View details →Convatec, EsteemBody Drainable Pouch REF:423653 10-35mm
Drainable large pouch may leak due to manufacturing issue.
Mar 25, 2026
View details →Product Name: ID NOW" Influenza A & B 2 Model/Catalog Number: 427-000 Software Version: Not Applicable Product Description: Influenza A/B virus nucleic acid IVD, kit, nucleic acid technique Component: Not Applicable
It was confirmed that the impacted lot has a higher occurrence of invalid rates when compared to the product Instructions for Use.
Mar 25, 2026
View details →Cub Pediatric Crib, Model FL19H
Cribs sold in the USA are missing two access door warning labels.
Mar 25, 2026
View details →Convatec, EsteemBody Drainable Pouch REF:423643, 10-45mm
Drainable large pouch may leak due to manufacturing issue.
Mar 25, 2026
View details →Centricity Universal Viewer Software Versions 6.0 through 6.0 Sp10.4.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-406, 2100169-001, 2088026-985, 2088026-971, 2088026-614, 2088026-713, 2088026-994, 2100491-006, 2088026-723, 2088026-308, 2104867-044, 2100491-012, 2088026-962, 2088026-969, 2102676-001, 2088026-936, 2088026-960, K1220JWED, 2088026-806, 2088026-506, 2088026-026, 2088026-970, 2088026-943, 2088026-990, 2088026-940, 2088026-959, 2088026-967, 2088026-989, 2104867-045, 2088026-993, 2088026-306, 2102675-010, 2088026-948, 2088026-957, 2088026-973, 2088026-946, 2088026-715, 2088026-906
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
Mar 25, 2026
View details →Brand Name: Leksell Gamma Knife Product Description: Leksell Gamma Knife¿ Perfexion Leksell Gamma Knife¿ Icon" Leksell Gamma Knife¿ Elekta Esprit Leksell Gamma Knife¿ (available models Elekta Esprit, Icon", and Perfexion") is a radiosurgery system for use in the stereotactic irradiation of head structures. Surgery is achieved by delivering a prescribed dose as one or more shots of ionizing radiation to the exact site of the target.
Due to adapter not being properly aligned or locked in place.
Mar 25, 2026
View details →Brand Name: STA Liatest D-Di Product Name: STA Liatest D-Di Model/Catalog Number: REF 00515 Product Description: The STA¿ - Liatest ¿ D-Di kit is an immuno-turbidimetric assay for the quantitative determination of D-dimer in venous plasma (in 3.2 % sodium citrate) for use on STA-R¿, STA Compact¿ and STA Satellite¿ analyzers by professional laboratory personnel. The STA¿ - Liatest¿ D-Di is intended for use in conjunction with a clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) and deep venous thrombosis (DVT) in outpatients suspected of PE or DVT. Component: No
After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
Mar 25, 2026
View details →MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
Mar 25, 2026
View details →Battery Charging Station; Model: 0998-00-0802;
The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
Mar 25, 2026
View details →CLARITY II Laser System; Model No. 1110200210.
Reports of devices sparking/popping and potentially burning patients.
Mar 18, 2026
View details →Endo-Model Replacement Plateau; Item Number: 15-0027/11;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →Owner's Booklets and Instructions for Use that are used with the
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
Mar 18, 2026
View details →Penner Pacific Bathing Spa, Model Numbers 360020-1P
The device does not bear a unique device identifier.
Mar 18, 2026
View details →Artelon FlexBand Dynamic Matrix Ref: 31057
Augmentation devices failed bacterial endotoxin testing.
Mar 18, 2026
View details →CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment
Mar 18, 2026
View details →Endo-Model Replacement Plateau; Item Number: 15-8030/12;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →GEM Premier 5000 PAK, Part No. 00055360004.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →GEM Premier 5000; Part No. 00055415005.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Raz-AT (Attendant Tilt) Mobile Shower Commode Chair, Catalog Number Z300 UDI-DI code: (01)00628758003007 Raz-ART (Attendant Rotational Tilt) Mobile Shower Commode Chair, Catalog Number Z333 UDI-DI code: (01)00628758003335 Raz-AT600 (Bariatric Tilt) Mobile Shower Commode Chair, Catalog Number Z360 UDI-DI code: (01)00628758003601 Raz-SP (Self Propel) Mobile Shower Commode Chair, Catalog Number Z200 UDI-DI code: (01)00628758001003 Raz-AP (Attendant Propel) Mobile Shower Commode Chair, Catalog Number Z100 UDI-DI code: (01)00628758001003 The Raz Mobile Shower Commode Chair is intended to be used by people with mobility impairments, who may have difficulty in standing or keeping their balance while completing activities such as showering, toileting, or personal hygiene.
Due to a modification by a vendor and may be reinstalled without properly engaging the outside rear seat brackets that are used to keep the seat in place.
Mar 18, 2026
View details →Owner's Booklets and Instructions for Use that are used with the
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
Mar 18, 2026
View details →GEM Premier 5000; Part No. 00055430008.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Medline Kits containing Cardinal Health s Aqua-Seal Chest Drainage Unit: 1) OPEN HEART, Model Number: DYNJ902487A; 2) CARDIAC PROCEDURE, Model Number: DYNJ908126; 3) CARDIO CONGENITO, Model Number: DYNJ910851
The instructions for use (IFUs) on specified Cardinal Health Chest Drainage Units (CDUs) and accessories have been updated to clarify the intended target patient population as adults, 18 years and older.
Mar 18, 2026
View details →Endo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/05;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →Endo-Model Replacement Plateau; Item Number: 15-8521/15;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →Brand Name: Equinoxe Ergo Product Name: Ergo Modular Impactor Handle Model/Catalog Number: 321-09-05
Impactor handle may be missing cross-pin
Mar 18, 2026
View details →DxC 700 AU, REF: B86444, B86446
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.
Mar 18, 2026
View details →GEM Premier 5000; Part No. 00055445008.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →GEM Premier 5000; Model No. 00055430010.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →GEM Premier 5000 PAK; Part No. 00055407511.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Penner Pacific Bathing Spa, Model Numbers 360030-1C, 360030-1CE, 360030-1CTS, 360030-XCE, 360030-XC, 360030-1CT, 360030-1C, 360030-XC, 360030-XCT, 360030-1CTS, 360030-XCTS, 3600301-CTS, 360030-XCE, 360030-1CT, 360030-1CE, 36030-XCT
The device does not bear a unique device identifier.
Mar 18, 2026
View details →Brand Name: APTUS Product Name: 2.5 TriLock Screw 16mm, HD7, 1/Pkg Model/Catalog Number: A-5750.16/1 Software Version: Not applicable. Product Description: The APTUS fixation systems are used for fractures, osteotomies and arthrodesis of the hand, forearm, shoulder and foot. Component: Not applicable.
Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws.
Mar 18, 2026
View details →GEM Premier 5000 PAK; Part No. 00055360008.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →GEM Premier 5000; Part No. 00055415008.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Penner Pacific Bathing Spa, Model Numbers 360020-1EP
The device does not bear a unique device identifier.
Mar 18, 2026
View details →PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Infant/Child Reduced Energy Defibrillation Electrodes are pre gelled, self-adhesive, therapy electrodes that allow hands free defibrillation. These electrodes reduce the energy delivered to the patient by a factor of 4:1. The Infant/child reduced energy defibrillation electrodes are therapy electrodes designed to allow personnel who are trained on AED device operation and in basic life support, or other physician authorized emergency medical response systems, to safely and effectively defibrillate infants and children who are less than 8 years old or weigh less than 25kg (55lbs).
Due to pediatric defibrillator electrode delamination
Mar 18, 2026
View details →GEM Premier 5000 PAK; Part No. 00055407510.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →GEM Premier 5000 PAK; Part No. 00055360011.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →GEM Premier 5000 PAK; Part No. 00055407508.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Owner's Booklets and Instructions for Use that are used with the
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
Mar 18, 2026
View details →MEDLINE UNNA-Z STRETCH ZINC PASTE BANDAGE, REF NONUNNAS40 KOB article #40307010
Due to labeling error. Incomplete ingredients listed. Labeling did not include castor oil or calamine.
Mar 18, 2026
View details →GEM Premier 5000; Part No. 00055445004.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Artelon FlexBand Plus Ref: 41054 & 41057
Augmentation devices failed bacterial endotoxin testing.
Mar 18, 2026
View details →Penner Pacific Bathing Spa, Model Numbers 360010-2WT, 360010-1, 370000-1W, 360010-1WT, 362010-1, 370000-1WL, 360010-1L, 360010-1W, 360010-XWTL, 370000-XWL, 360010-1WTL, 360010-XWT, 360010-2, 370000-2W, 360010-2L, 370000-2WL, 360010-X, 370000-XW, 360010-2W, 362010-2, 360010, 37000-XWL, 362010-X, 362010-XL, 362010-1L, 360010-XL, 37000-1W, 360010-XW, 3700000-1W, 362010-2L, 36010-XWT, 370000-1L, 360010-2WT, 360000-1, 360010-XWSB, 360010-XLSB
The device does not bear a unique device identifier.
Mar 18, 2026
View details →GEM Premier 5000; Part No: 00055415011.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Owner's Booklets and Instructions for Use that are used with the
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
Mar 18, 2026
View details →GEM Premier 5000; Part No. 00055415010.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →GEM Premier 5000; Part No. 00055430004.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Penner Pacific Bathing Spa, Model Numbers 361910-1A, 361910-1A, 361910-XA, 361910-2I, 361910-1A, 361910-1, 361910-XI, 361910-1I, 361910-2A, 36191-1A, 361910-X
The device does not bear a unique device identifier.
Mar 18, 2026
View details →Penner Pacific Bathing Spa, Model Numbers 560010-2L, 560010-XL, 560010-1L, 560010-XL
The device does not bear a unique device identifier.
Mar 18, 2026
View details →Brand Name: Equinoxe Product Name: Equinoxe Core Instrument Kit Model/Catalog Number: KIT-311X
Impactor handle may be missing cross-pin
Mar 18, 2026
View details →Artelon FLEXBAND TWIST .12 Ref: TW012 3.85x17mm HEX Anchor (Qty 2) 5.0x17mm HEX Anchor (Qty 1) 0.5x12cm FLEXBAND (Qty 1)
Augmentation devices failed bacterial endotoxin testing.
Mar 18, 2026
View details →Penner Pacific Bathing Spa, Model Numbers 760010-1, 770010-1
The device does not bear a unique device identifier.
Mar 18, 2026
View details →Endo-Model Replacement Plateau; Item Number: 15-0027/15;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →Clearest Strep-A Cassette Test. Test to determine the presence of group A streptococci (Strep A) in throat swab samples.
Product not cleared by the FDA.
Mar 18, 2026
View details →Penner Pacific Bathing Spa, Model Numbers 350010-1, 350000-2, 350010-X
The device does not bear a unique device identifier.
Mar 18, 2026
View details →Medline Medcrest Surgical Gowns: MDTGXC4J5XL DBQ-GOWN,SURG,XALT,L4, CRTCL CVR, GRN, 5 MDTGXC4JL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,L MDTGXC4JXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XL MDTGXC4JXXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XX MDTGXP4JL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,L MDTGXP4JXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTGXP4JXXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XX MDTL1002862EF DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTZ1000893AJ DBD-GOWN,SRG,PNL CVR, GREEN,SNPS,XL
Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.
Mar 18, 2026
View details →HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03 Model / UDI-DI codes: Vital Sync 3.3.1 / 10884521844933 Vital Sync 3.4.0 / 10884521844940 Vital Sync 3.4.1 / 10884521847569 (OUS Only) Vital Sync 3.4.2 / 10884521856127 Product Number: VSLBASE04P / VITALSYNCSW04 Model / UDI-DI codes: Vital Sync 4.0.0 / 10884521849358 Vital Sync 4.0.2 / 10884521856134 The Vital Sync" Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Virtual Patient Monitoring Platform, 3rd Party Annunciation Systems, Electronic Medical Record (EMR) and Clinical Information System (CIS).
Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.
Mar 18, 2026
View details →Penner Pacific Bathing Spa, Model Numbers 860010-1L
The device does not bear a unique device identifier.
Mar 18, 2026
View details →10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case.
During the bulk packaging process, some Luer Slip syringes were incorrectly included in a batch designated for Luer Lok syringes.
Mar 18, 2026
View details →IBA Proton Therapy System - PROTEUS 235
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
Mar 18, 2026
View details →Endo-Model Replacement Plateau; Item Number: 15-2835/12;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →Endo-Model Replacement Plateau; Item Number: 15-2836/11;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →GEM Premier 5000 PAK; Part No. 00055415004.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Brand Name: Bisaf Strep A Self-Test. Test for detection of group A streptococcal antigens in throat swabs.
Product not cleared by the FDA.
Mar 18, 2026
View details →GEM Premier 5000; Part No. 00055430011.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Medline Surgical Drapes: MDTBTCS6070GN DBD-DRAPE,COVER,TABLE,RESIS/XALT GRN,60X MDTBTCS6090GN DBD-DRAPE,COVER,TABLE,RESIS/XALT GRN,60X MDTDXUNDBGN DBD-DRAPE,UNDERBUTTOCK,28 X 50.5,XALT GR MDTMAYOXGN DBD-DRAPE,COVER,MAYO STAND,RESISTAT/XALT MDTZ1003656GN DBD-POCKET SHEET,RESISTAT/XALT
Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.
Mar 18, 2026
View details →GEM Premier 5000; Part No. 00055445010 & 00055445011.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Penner Pacific Bathing Spa, Model Numbers 390010-1, 390010-X, 390010-2
The device does not bear a unique device identifier.
Mar 18, 2026
View details →Endo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/07;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →GEM Premier 5000 PAK; Part No. 00055360010.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Brand Name: APTUS Product Name: 2.8 TriLock Screw 16mm, HD7, 1/Pkg Model/Catalog Number: A-5850.16/1 Software Version: Not applicable. Product Description: The APTUS fixation systems are used for fractures, osteotomies and arthrodesis of the hand, forearm, shoulder and foot. Component: Not applicable.
Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws.
Mar 18, 2026
View details →Wondfo Streptococcal A At-Home Self-Test. At-home test for group A beta-hemolytic streptococcal infection.
Product not cleared by the FDA.
Mar 18, 2026
View details →Endo-Model Replacement Plateau; Item Number: 15-0027/12;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →Endo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/02;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →Endo-Model Replacement Plateau; Item Number: 15-8521/09;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →Endo-Model Replacement Plateau; Item Number: 15-8521/11;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000;
Recall of kits which utilized ChloraPrep Triple Swabsticks from BD. Open or compromised seals on the sterile swabstick packaging, which could result in loss of sterility or drying of the antiseptic solution. This component is placed outside of the sterile barrier of the tray manufactured by Boston Scientific, which contains the majority of the kit components. As such, inclusion of the affected ChloraPrep Triple Swabsticks does not degrade the integrity of the other components within the PEG kit.
Mar 18, 2026
View details →Endo-Model Replacement Plateau; Item Number: 15-0027/16;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Mar 18, 2026
View details →GEM Premier 5000 PAK; Part No. 00055407504.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Mar 18, 2026
View details →Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure
Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.
Mar 4, 2026
View details →Brand Name: Hitachi Proton Beam Therapy REF: PROBEAT-CR
Software anomaly in the patient positioning system may result in positional discrepancy.
Mar 4, 2026
View details →Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE INSERTION SKU DYNJ63347A
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Mar 4, 2026
View details →PIE Trolley System Model: 2005
Lack of 510K clearance
Mar 4, 2026
View details →Olympus Inner Sheath, Long; Model No. WA22017T. Inner sheath for urological and gynecological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Medline Convenience Kits: 1) NAIL KIT, Model Number: POD14214
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
Mar 4, 2026
View details →Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Mar 4, 2026
View details →Olympus Resection Sheath, 24 Fr., with Deflecting Obturator; Model No. A22041T. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 Medline DIALYSIS CHANGE KIT SKU EBSI1453A Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Mar 4, 2026
View details →Olympus Resection Sheath, 28 Fr. with Deflecting Obturator. Model No. A22043T. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Madsen® AccuScreen® TEOAE/DPOAE/ABR Probe Recall
Recalls and safety alerts
Mar 4, 2026
View details →Olympus Sheath, 10 Fr. Model No. A37004A. for endoscopic diagnosis and treatment in urological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →MRI LED Mobile Exam Light Battery REF SL-111 This non-magnetic UL and CE approved battery is used in conjunction with SL-110, Mobile Exam Light
Due to battery component overheating while charging resulting in melting of internal components and causing smoke.
Mar 4, 2026
View details →Olympus Resection Sheath, 28 Fr.; Model No. A22043A. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Olympus Inner Sheath; Model No. A2660. for endoscopic diagnosis and treatment in urological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Olympus Resection Sheath, 26 Fr. with Deflecting Obturator; Model No. A22042T. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Mar 4, 2026
View details →Brand Name: Reverse Shoulder Prosthesis (RSP) Product Name: RSP HUMERAL SOCKET INSERT, 32MM +4MM, STANDARD HXe-plus Model/Catalog Number: 509-00-432
Knee and Humeral socket implants contain incorrect labeling.
Mar 4, 2026
View details →Olympus Inner Sheath, Long; Model No. WA22017A. Inner sheath for urological and gynecological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Olympus Inner Sheath. Model No. A2641. for endoscopic diagnosis and treatment in urological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Olympus Resection Inner Sheath, for 26 Fr. Outer Sheath. Model No. A22040A. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Medline Convenience Kits: 1) NAIL KIT, Model Number: DYKM1528
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
Mar 4, 2026
View details →Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1
During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
Mar 4, 2026
View details →Medline Kits containing Tego Connectors: Medline IR Pack, SKU DYNJ67205C Medline NO CATHETER DIALYSIS INSERTION, SKU CVI5200
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Mar 4, 2026
View details →Brand Name: Reverse Shoulder Prosthesis (RSP) Product Name: RSP STANDARD HUMERAL SOCKET INSERT, 36MM, HXe-plus Model/Catalog Number: 509-00-036
Knee and Humeral socket implants contain incorrect labeling.
Mar 4, 2026
View details →Olympus Inner Sheath. Model No. A4741. for endoscopic diagnosis and treatment in gynecological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 EBSI1596 EBSI1781 EBSI1812 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Mar 4, 2026
View details →Olympus Resection Sheath; Model No. A2666. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Meijer STERILE saline wound wash, First Aid Cleansing Spray, 0.9% USP SODIUM CHLORIDE, 7.1FL OZ (210 mL), DBD-CAN,SPRAY,7.1-OZ,SALINE,WOUND WASH,S, Model Number MJSALINE7
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
Mar 4, 2026
View details →Olympus Resection Sheath. Model No. A2666T. for endoscopic diagnosis and treatment in urological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Olympus Resection Sheath, 8 mm. Model No. A42011A. Resection sheath for gynecological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Brand Name: Hitachi Proton Beam Therapy System Ref: PROBEAT-FR
Software anomaly in the patient positioning system may result in positional discrepancy.
Mar 4, 2026
View details →Olympus Resection Inner Sheath for 26 Fr. Outer Sheath. Model No. A22040T. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →PIE PAK Models: P2HC-A, P2HC-S, P2HC
Lack of 510K clearance
Mar 4, 2026
View details →Olympus Resection Sheath, 24 Fr; Model No. A22041A. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Olympus Cystoscope Outer Sheath; Model/Catalog number: WA22810A;
Product being removed due to incompatibility when used in conjunction with a GreenLight Laser for BPH therapy which could damage the tip of the device.
Mar 4, 2026
View details →Medline Convenience Kits: 1) TRUNK KIT W/EXPIRATION, Model Number: DYKM1361A; 2) TRUNK KIT 1EA, Model Number: DYKM2013; 3) TRUNK KIT 1EA, Model Number: DYKM2013A; 4) RN TRUNK KIT, Model Number: DYKM2699; 5) TRUNK KIT, Model Number: DYKTRUNK1
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
Mar 4, 2026
View details →Olympus Inner Sheath 21 Fr. Model No. A2660T. for endoscopic diagnosis and treatment in urological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Olympus Resection Sheath, 26 Fr.; Model No. A22042A. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Medline Sterile Saline Wound Wash, Cleansing & Wound Irrigation, USP Normal Saline (0.9% Sodium Chloride), 7.1 FL OZ (210 mL), DBD-CAN,SPRAY,7.1-OZ,SALINE,WOUND WASH,S, Model Number MDSALINE7
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
Mar 4, 2026
View details →Olympus Inner Sheath. Model No. A2642. for endoscopic diagnosis and treatment in urological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Campy CVA Medium 100/PK, R01272
Customer complaints report low to no recovery of Campylobacter Jejuni ATCC 33291 on identified lot of medium.
Mar 4, 2026
View details →Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 7L 16MM, VE Model/Catalog Number: 341-16-707
Knee and Humeral socket implants contain incorrect labeling.
Mar 4, 2026
View details →Olympus Resection Sheath. Model No. A22014A and A22014T. for endoscopic diagnosis and treatment in urological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Mar 4, 2026
View details →Vasopressin Recall
Recalls and safety alerts
Mar 4, 2026
View details →Sterile Contro-Vac Whistle Tip Suction Cath W/2 Gl... Recall
Recalls and safety alerts
Feb 27, 2026
View details →Burlington Medical Thyroid Shields Recall
Recalls and safety alerts
Feb 27, 2026
View details →Paradigm and MiniMed 600 and 700 series pumps Recall
Recalls and safety alerts
Feb 26, 2026
View details →Smartmix Tips Combi Recall
Recalls and safety alerts
Feb 26, 2026
View details →Champignons et Adaptogènes: Affected lot contains ... Recall
Recalls and safety alerts
Feb 26, 2026
View details →Olympus Thunderbeat, 5mm, 35 cm, Pistol Grip
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Olympus Thunderbeat 5 mm, 20 cm, Front-Actuated Grip
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Medline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number: DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number: DYNJ56436B; 3) CPNB SETUP TRAY, Model Number: DYNJRA1979; 4) CPNB SETUP TRAY, Model Number: DYNJRA1979A; 5) CPNB SETUP TRAY, Model Number: SAMPA0108
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Olympus Thunderbeat, 5 mm, 45 cm Inline Grip
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Olympus Thunderbeat, 5 mm, 35 cm, Inline Grip
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Brand Name: Prodisc Product Name: Prodisc C SK Model/Catalog Number: PDSL6 Software Version: N/A Product Description: Total cervical disc replacement Component: No
Products were mislabeled as the 6mm product but included the 5 mm product.
Feb 25, 2026
View details →Olympus Thunderbeat, 5 mm, 45 cm Front-Actuated Grip
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Medline Convenience Kits: 1) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326; 2) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326A; 3) DRAWER #5, Model Number: ACC010475; 4) DRAWER 1 NEURO CART, Model Number: ACC010542; 5) DRAWER 2 NEURO CART, Model Number: ACC010543; 6) DRAWER 4, Model Number: ACC010722; 7) MALIGNANT HYPERTHERMIA CART, Model Number: ACC010896; 8) LIVER TRANSPLANT #57 CDS-2, Model Number: CDS860063U; 9) THOROCOTOMY CDS #24-RF, Model Number: CDS860066T; 10) CYSTO, Model Number: CDS981795G; 11) MAJOR PROCEDURE CDS, Model Number: CDS982414N; 12) TOTAL KNEE CDS, Model Number: CDS982823R; 13) VASCULAR CDS-1, Model Number: CDS982895I; 14) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069F; 15) HEART CABG CDS, Model Number: CDS983376S; 16) CRANIOTOMY, Model Number: CDS983611N; 17) CUH KIDNEY TRANSPLANT CDS, Model Number: CDS983653K; 18) CUH KIDNEY TRANSPLANT CDS, Model Number: CDS983653L; 19) CUH KIDNEY TRANSPLANT CDS, Model Number: CDS983653M; 20) MAJOR ABDOMINAL CDS, Model Number: CDS983908Q; 21) NHP ROBOTIC GENERAL CDS, Model Number: CDS984244I; 22) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431M; 23) MAJOR ABDOMINAL CDS (NO GOWNS), Model Number: CDS985557L; 24) PREP KIT, Model Number: DYK1022113CP; 25) BARRIER KIT,UNIV OF ALABAMA, Model Number: DYK1060193B; 26) PERSONAL ITEM KIT, Model Number: DYKA1131; 27) GI PACK, Model Number: DYKE1441D; 28) GI PACK, Model Number: DYKE1441G; 29) BAPTIST FLOYD BRONCHOSCOPY, Model Number: DYKE1456C; 30) VIDEO BRONCHOSCOPY TRAY, Model Number: DYKE1666; 31) GI LAB OTHER ENDO KIT, Model Number: DYKE1721A; 32) MATERNITY KIT, Model Number: DYKL1343; 33) BUNNY KIT, Model Number: DYKM1054C; 34) ASST STERILE PPE, Model Number: DYKM1270; 35) MASK CAP AND GOWN KIT, Model Number: DYKM1427A; 36) BLOOD PRECAUTION KIT, Model Number: DYKM1687; 37) HELPERS KIT, Model Number: DYKM1832; 38) BRONCHOSPY SUPPLY KIT, Model Number: DYKM1898; 39) OP UROLOGY KIT, Model Number: DYKM2316; 40) ICU PACK, Model Number: DYKM2671; 41) KIT OMF LEFORTE, Model Number: DYKMBNDL1; 42) KIT GEN SURG RECTAL, Model Number: DYKMBNDL103; 43) KIT GEN SURG RECTAL, Model Number: DYKMBNDL103A; 44) KIT GEN SURG RECTAL, Model Number: DYKMBNDL103B; 45) KIT UROLOGY EXPLORATORY, Model Number: DYKMBNDL115; 46) KIT UROLOGY EXPLORATORY, Model Number: DYKMBNDL115A; 47) KIT UROLOGY EXPLORATORY, Model Number: DYKMBNDL115B; 48) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116C; 49) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117B; 50) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121D; 51) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121G; 52) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121I; 53) KIT ACES INSERT PEG, Model Number: DYKMBNDL130B; 54) KIT OMF PEDS MINOR, Model Number: DYKMBNDL139; 55) KIT THOR ESOPH PERF.REPAIR, Model Number: DYKMBNDL140; 56) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142A; 57) KIT OMF TMJ ARTHROPLASTY (OPEN, Model Number: DYKMBNDL146A; 58) KIT OMF TMJ ARTHROPLASTY (OPEN, Model Number: DYKMBNDL146C; 59) KIT OMF TMJ ARTHROPLASTY (OPEN, Model Number: DYKMBNDL146D; 60) KIT THORACIC THYMECTOMY, Model Number: DYKMBNDL161; 61) KIT ORTHO TOTAL SHOULDER, Model Number: DYKMBNDL162B; 62) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165B; 63) KIT OMF MINOR, Model Number: DYKMBNDL166; 64) KIT TRANSPLANT PARTIAL LIVER, Model Number: DYKMBNDL167; 65) KIT TRANSPLANT PARTIAL LIVER, Model Number: DYKMBNDL167A; 66) KIT TRANSPLANT LIVER EX-LAP, Model Number: DYKMBNDL167B; 67) KIT LAP DONOR NEPHRECTOMY, Model
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, Model Number: CDS980833Q; 7) TOTAL KNEE CDS-LF, Model Number: CDS980833R; 8) TOTAL HIP PROCEDURE, Model Number: CDS980865V; 9) TOTAL HIP PROCEDURE, Model Number: CDS980865W; 10) TOTAL HIP PROCEDURE, Model Number: CDS980865X; 11) TOTAL HIP PROCEDURE, Model Number: CDS980865Y; 12) TOTAL KNEE CDS, Model Number: CDS981454K; 13) TOTAL KNEE CDS, Model Number: CDS981454L; 14) KNEE ARTHROSCOPY CDS, Model Number: CDS981502K; 15) TOTAL KNEE, Model Number: CDS981505O; 16) TOTAL KNEE, Model Number: CDS981505P; 17) TOTAL KNEE, Model Number: CDS981505Q; 18) TOTAL KNEE, Model Number: CDS981505R; 19) OR LAMINECTOMY CDS, Model Number: CDS981858F; 20) LAMI FUSION, Model Number: CDS982035W; 21) ARMC TOTAL KNEE CDS, Model Number: CDS982281O; 22) ARMC TOTAL KNEE CDS, Model Number: CDS982281P; 23) ARMC TOTAL KNEE CDS, Model Number: CDS982281S; 24) TOTAL HIP CDS, Model Number: CDS983036K; 25) TOTAL KNEE CDS, Model Number: CDS983042K; 26) LOWER EXTREMITY CDS, Model Number: CDS983049K; 27) ZALE SPINE CDS, Model Number: CDS983182G; 28) TOTAL KNEE CDS, Model Number: CDS983297M; 29) SITTING CERVICAL CDS, Model Number: CDS983480G; 30) SHOULDER CDS, Model Number: CDS983581K; 31) SHOULDER ARTHROSCOPY CDS, Model Number: CDS983838G; 32) TOTAL HIP CDS, Model Number: CDS983840I; 33) TOTAL JOINT CDS-LF, Model Number: CDS983863D; 34) SHOULDER ARTHROSCOPY CDS, Model Number: CDS983879C; 35) TOTAL JOINT, Model Number: CDS983903D; 36) GB TOTAL KNEE, Model Number: CDS984147K; 37) GB LOWER EXTREMITY, Model Number: CDS984156L; 38) LOWER EXTREMTIY, Model Number: CDS984254K; 39) LOWER EXTREMTIY, Model Number: CDS984254L; 40) LOWER EXTREMTIY, Model Number: CDS984254M; 41) LAMI MICRODISC-LF, Model Number: CDS984527I; 42) EXTREMITY CDS-LF, Model Number: CDS984607G; 43) EXTREMITY CDS-LF, Model Number: CDS984607I; 44) HANA/FRACTURE TABLE CDS-LF, Model Number: CDS984610C; 45) TOTAL JOINT CDS, Model Number: CDS985202B; 46) TOTAL JOINT CDS, Model Number: CDS985202C; 47) TRAUMA-LF, Model Number: CDS985279K; 48) ARTHROSCOPY CDS, Model Number: CDS985316A; 49) EXTREMITY CDS, Model Number: CDS985317C; 50) EXTREMITY CDS, Model Number: CDS985317D; 51) EXTREMITY CDS, Model Number: CDS985317F; 52) EXTREMITY CDS, Model Number: CDS985317G; 53) KNEE ARTHROSCOPY CDS, Model Number: CDS985425J; 54) TOTAL HIP PACK (CDSOPM834)-LF, Model Number: CDSOPM834W; 55) JOINT ASPIRATION KIT, Model Number: DYKM1161; 56) TOTAL KNEE SUPPLEMENT PACK NS, Model Number: DYKM2106; 57) KIT ORTHRO TOTAL HIP DR. NANDI, Model Number: DYKMBNDL137; 58) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138A; 59) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141; 60) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141A; 61) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141B; 62) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142C; 63) KIT NEURO CRANIOTOMY ACOUSTIC, Model Number: DYKMBNDL143; 64) KIT NEURO CRANIOTOMY FOR TUMOR, Model Number: DYKMBNDL144; 65
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) OR ARTERIOGRAM, Model Number: CDS985094F; 2) KIT SURG ONC LAP DIAG/CV CATH, Model Number: DYKMBNDL153; 3) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4A; 4) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4B; 5) VEIN PROCEDURE KIT, Model Number: DYNDA2076C; 6) CARDIAC CATH PACK-LF, Model Number: DYNJ0150605AC; 7) CATH LAB PACK-LF, Model Number: DYNJ0648473Y; 8) NBMC CATH PACK-LF, Model Number: DYNJ0803367I; 9) ANGIO PACK-LF, Model Number: DYNJ0954970I; 10) CARDIAC CATH PACK, Model Number: DYNJ19946M; 11) CATH LAB PACK, Model Number: DYNJ30955I; 12) CATH LAB PACK, Model Number: DYNJ31773K; 13) CATH PACK-LF, Model Number: DYNJ33061D; 14) DB CCL OAKWOOD PACK-LF, Model Number: DYNJ33620D; 15) CARDIAC CATH PACK MRH-LF, Model Number: DYNJ35300M; 16) CATH LAB PK, Model Number: DYNJ35643C; 17) CARDIAC CATH PACK-LF, Model Number: DYNJ39066B; 18) CATH PACK-LF, Model Number: DYNJ43101C; 19) OR ANGIO PACK-LF, Model Number: DYNJ43415D; 20) OR ANGIO PACK-LF, Model Number: DYNJ43415F; 21) ANGIOPLASTY PACK, Model Number: DYNJ44700J; 22) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJ44850J; 23) RADIOLOGICAL PACK, Model Number: DYNJ44889O; 24) CARDIAC CATH PACK, Model Number: DYNJ47370B; 25) ANGIOGRAPHY PACK, Model Number: DYNJ53129D; 26) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJ56772G; 27) DR. MOHINDRA PACK, Model Number: DYNJ57602G; 28) NE ENDO VASCULAR PACK, Model Number: DYNJ57940B; 29) CARDIAC CATH PACK - Q1, Model Number: DYNJ58048C; 30) OR ANGIO, Model Number: DYNJ59931B; 31) LPCH PERRY CATH PACK, Model Number: DYNJ67378C; 32) HOSPITAL REGIONAL DE CONCEPCIO, Model Number: DYNJ67918A; 33) CATH LAB PACK, Model Number: DYNJ68633; 34) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197B; 35) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197C; 36) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197D; 37) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197F; 38) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197G; 39) BASIC CARDIAC CATH PACK, Model Number: DYNJ81593B; 40) UMC EL PASO CATH PACK, Model Number: DYNJ83089; 41) DR. MOHINDRA PACK, Model Number: DYNJ83819A; 42) VZ VAS/ANG PACK, Model Number: DYNJ83981; 43) WCPN CARDIOLOGY PACK, Model Number: DYNJ84539; 44) CATH LAB, Model Number: DYNJ85511; 45) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJT1520K; 46) ANGIOGRAPHY TRAY, Model Number: MNS9130; 47) CVC BUNDLE KIT, Model Number: ECVC6225A; 48) CVC 3L 7F 20CM VANTEX, Model Number: ECVC7265A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Olympus Thunderbeat, 5 mm, 35 cm, Front-Actuated Grip Type S
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Medline Convenience Kits: 1) CHEST/BREAST-LF, Model Number: DYNJ83950
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) KIT UROLOGY SLING, Model Number: DYKMBNDL106; 2) KIT UROLOGY SLING, Model Number: DYKMBNDL106A; 3) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188; 4) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189; 5) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) OPEN HEART CDS, Model Number: CDS840015AQ; 2) OPEN HEART CDS, Model Number: CDS840023T; 3) OFF PUMP CABG CDS, Model Number: CDS840087AI; 4) OFF PUMP CABG CDS, Model Number: CDS840087AK; 5) OFF PUMP CABG CDS, Model Number: CDS840087AL; 6) OPEN HEART, Model Number: CDS840150S; 7) OPEN HEART, Model Number: CDS840150T; 8) OPEN HEART, Model Number: CDS840150U; 9) OPEN HEART, Model Number: CDS840150V; 10) OPEN HEART, Model Number: CDS840150W; 11) UNIVERSAL HEART CDS, Model Number: CDS840246O; 12) GENERAL AAA #11-RF, Model Number: CDS840261AB; 13) OPEN HEART ADULT, Model Number: CDS840396AA; 14) OPEN HEART ADULT, Model Number: CDS840396AB; 15) OPEN HEART ADULT, Model Number: CDS840396X; 16) OPEN HEART ADULT, Model Number: CDS840396Y; 17) OPEN HEART CDS, Model Number: CDS840428K; 18) OPEN HEART CDS, Model Number: CDS840428L; 19) PACEMAKER CDS, Model Number: CDS980839C; 20) KIT, CVC ADD A PACK HUM-LF, Model Number: CDS982564C; 21) OPEN HEART ADULT HUM, Model Number: CDS982566K; 22) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575K; 23) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575M; 24) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575O; 25) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575P; 26) OPEN HEART, Model Number: CDS983348S; 27) HEART CABG CDS, Model Number: CDS983376Q; 28) HEART CABG CDS, Model Number: CDS983376T; 29) HEART CABG CDS, Model Number: CDS983376U; 30) HEART CABG CDS, Model Number: CDS983376V; 31) OPEN HEART CDS, Model Number: CDS983513F; 32) OPEN HEART ADULT PART 2 CDS, Model Number: CDS983786D; 33) OPEN HEART ADULT PART 2 CDS, Model Number: CDS983786F; 34) OPEN HEART ADULT PART 2 CDS, Model Number: CDS983786G; 35) PEDS CARDIAC CDS, Model Number: CDS984117F; 36) OPEN HEART CDS, Model Number: CDS984276P; 37) OPEN HEART ACCESSORY CDS, Model Number: CDS984289J; 38) OPEN HEART ACCESSORY CDS, Model Number: CDS984289K; 39) OPEN HEART ACCESSORY CDS, Model Number: CDS984289L; 40) OPEN HEART ACCESSORY CDS, Model Number: CDS984289M; 41) OPEN HEART ACCESSORY CDS, Model Number: CDS984289N; 42) OPEN HEART ACCESSORY CDS, Model Number: CDS984289O; 43) OPEN HEART CDS, Model Number: CDS984355F; 44) OPEN HEART CDS, Model Number: CDS984355G; 45) OPEN HEART CDS, Model Number: CDS984355I; 46) OPEN HEART CDS, Model Number: CDS984355J; 47) OPEN HEART A, Model Number: CDS985105M; 48) OPEN HEART A, Model Number: CDS985105N; 49) OPEN HEART A, Model Number: CDS985105O; 50) VASCULAR CDS, Model Number: CDS985215C; 51) BASE HEART PACK-LF, Model Number: CVT159AA; 52) ENDO KIT, Model Number: DYKE1580B; 53) KIT VASCULAR MINOR, Model Number: DYKMBNDL11; 54) KIT VASCULAR CAROTID, Model Number: DYKMBNDL118; 55) KIT VASCULAR CAROTID, Model Number: DYKMBNDL118A; 56) KIT VASCULAR BYPASS GRAFT GENE, Model Number: DYKMBNDL120; 57) KIT VASCULAR PERCUTANEOUS TEVA, Model Number: DYKMBNDL14; 58) KIT VASCULAR PERCUTANEOUS TEVA, Model Number: DYKMBNDL14A; 59) KIT VASCULAR VNS, Model Number: DYKMBNDL152; 60) KIT VASCULAR VNS, Model Number: DYKMBNDL152A; 61) KIT THORACIC SYMPHATECTOMY, Model Number: DYKMBNDL187; 62) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189B; 63) KIT VASCULAR ENDOVENOUS ABLATI, Model Number: DYKMBNDL191; 64) KIT VASCULAR ANGIOGRAM GENERIC, Model Number: DYKMBNDL192; 65) KIT VASCULAR ENDOVENOUS ABLATI, Model Number: DYKMBNDL203; 66) KIT VASCULAR MEDIAN ARCUATE, Model Number: DYKMBNDL209A; 67) KIT VASCULAR MEDIAN ARCUATE, Model Number: DYKMBNDL209B; 68) CARDIAC ADULT CABG, Model Number: DYKMBNDL20A; 69) CARDIAC ADULT CABG, Model Number: DYKMBNDL20B; 70) KIT CARDIAC CSICU ADULT ECMO, Model Number: DYKMBNDL21; 71) KIT VASCULAR AAA REPAIR, Model Number: DYKMBNDL23A; 72) KIT CARDIAC ANEURYSM, Model Number: DYKMBNDL27A; 73) KIT CARDIAC ANEURYSM, Model Number: DYKMBNDL27B; 74) KIT PEDS CARDIAC ADULT, Model Number: DYKMBNDL4
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Surgical Drapes: 1) SHEET,DRAPE,40X58,STERILE, Model Number: DYNJP2410; 2) DRAPE,SHEET,ECONOMY,40X58",ST,20/CS, Model Number: DYNJP2411; 3) SHEET,DRAPE, 40X70,STERILE, Model Number: DYNJP2412; 4) DRAPE,MEDIUM,TAPE,40X70",ST,50/CS, Model Number: DYNJP2412T; 5) SHEET,DRAPE,53X77,STERILE, Model Number: DYNJP2414; 6) DRAPE,SHEET,3/4,53X77",ST,20/CS, Model Number: DYNJP2414R; 7) DRAPE,SHEET,ULTRAGARD,20/CS, Model Number: DYNJP2414UG; 8) DRAPE,RNF,53X77",ST,20/CS, Model Number: DYNJP2416; 9) SHEET,DRAPE,70X85,STERILE, Model Number: DYNJP2417; 10) DRAPE,FILM SHEET,44X65",ST,20/CS, Model Number: DYNJP2418; 11) DRAPE,SHEET,70X100",ST,12/CS, Model Number: DYNJP2419; 12) LEGGINGS,33X49",XL,PAIR,ST,CLR,20/CS, Model Number: DYNJP2462; 13) DRAPE,CHEST,FENESTRATION,15X10",ST,12/CS, Model Number: DYNJP2491; 14) DRAPE,MINOR PROC,6X6" FEN, ST,10/CS, Model Number: DYNJP2492; 15) SHEET, DRAPE, SPLIT, ST,10/CS, Model Number: DYNJP2498; 16) DRAPE,LAPAROTOMY,T-SHEET,ST,12/CS, Model Number: DYNJP3003; 17) DRAPE,LAPAROTOMY,T,ULTRAGARD,12/CS, Model Number: DYNJP3003UG; 18) DRAPE,LAPAROTOMY,POUCH,ST,12/CS, Model Number: DYNJP3008; 19) DRAPE,LAPAROTOMY,POUCH ULTRAGARD,12/CS, Model Number: DYNJP3008UG; 20) DRAPE,LAPAROTOMY,T,PEDIATRIC,ST,10/CS, Model Number: DYNJP3009; 21) DRAPE,LAPAROTOMY,ABDOMINAL,ST,12/CS, Model Number: DYNJP3101; 22) DRAPE,LAPROSCOPY,CHOLE,W/TROUGH,ST,8/CS, Model Number: DYNJP3102A; 23) DRAPE,LAP,CHOLE,W/TROUGH ULTRAGARD,12/CS, Model Number: DYNJP3102UG; 24) DRAPE,ABDOMINAL,MAJOR,ST,8/CS, Model Number: DYNJP3103; 25) DRAPE,ABDOMINAL,MAJOR,PCH/TRGH,ST,8/CS, Model Number: DYNJP3109; 26) DRAPE,TOP,102X53",ST,14/CS, Model Number: DYNJP4002; 27) DRAPE,CARDIOVASCULAR,SPLIT,ST,8/CS, Model Number: DYNJP4003; 28) DRAPE,ANGIO,BRACH,ECLIPSE,38X44,ST,40/CS, Model Number: DYNJP4106; 29) DRAPE,FEMORAL ANGIOGRAPHY,W/POUCH,8/CS, Model Number: DYNJP4119; 30) DRAPE,TOP,ARMCOVERS,106X59,STERILE,20/CS, Model Number: DYNJP4215; 31) DRAPE,UNDERBUTTOCK,PCH,ST,20/CS, Model Number: DYNJP6002; 32) DRAPE,UNDERBUTTOCK,GRAD POUCH,PORT,20/CS, Model Number: DYNJP6006; 33) DRAPE,EENT,SPLIT,ST,8/CS, Model Number: DYNJP7001; 34) SHEET,T,THYROID,ST,12/CS, Model Number: DYNJP7003; 35) DRAPE,EXTREMITY,89X128",ST,12/CS, Model Number: DYNJP8002; 36) DRAPE,EXTREMITY,ULTRAGARD,12/CS, Model Number: DYNJP8002UG; 37) T-DRAPE,EXTREMITY,ST,6/CS, Model Number: DYNJP8003; 38) T-DRAPE,EXTREMITY,ULTRAGARD,6/CS, Model Number: DYNJP8003UG; 39) DRAPE,HAND,ST,18/CS, Model Number: DYNJP8005; 40) DRAPE,HAND,ULTRAGARD,18/CS, Model Number: DYNJP8005UG; 41) DRAPE,HIP,W/POUCH,ST,5/CS, Model Number: DYNJP8201; 42) DBD-DRAPE,HIP,W/POUCHES, ULTRAGARD, Model Number: DYNJP8201UG; 43) DRAPE,ORTHOMAX,HIP,W/POUCH,6/CS, Model Number: DYNJP8211; 44) SHEET,ORTHO,SPLIT,ST,10/CS, Model Number: DYNJP8301; 45) DRAPE,SPLIT,77X120",10/CS, Model Number: DYNJP8304A; 46) ULTRAGARD SPLIT DRAPE,10/CS, Model Number: DYNJP8304UG; 47) DRAPE,SHOULDER,BEACH CHAIR,ST,5/CS, Model Number: DYNJP8412; 48) DRAPE,SHOULDER,BEACH CH,ULTRAGARD,5/CS, Model Number: DYNJP8412UG; 49) DRAPE,SHOULDER,BEACH CHAIR,ST,5/CS, Model Number: DYNJP8414; 50) DRAPE,LITHOTOMY,ST,8/CS, Model Number: DYNJP9001; 51) DRAPE,LAVH,ST,8/CS, Model Number: DYNJP9103; 52) DRAPE,VETERINARY,CUSTOM,SM,20/CS, Model Number: MED901; 53) DRAPE,VETERINARY,CUSTOM,M,20/CS, Model Number: MED902; 54) DRAPE,VETERINARY,CUSTOM,L,20/CS, Model Number: MED903
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) C-SECTION PACK, Model Number: CDS820080N; 2) C-SECTION CDS, Model Number: CDS983523G; 3) C-SECTION, Model Number: CDSCCR611O; 4) C-SECTION, Model Number: CDSCCR611P; 5) LUD C-SECTION-LF, Model Number: DYNJ0101616I; 6) C-SECTION TRAY, Model Number: DYNJ15494G; 7) C-SECTION TRAY-LF, Model Number: DYNJ19529P; 8) C-SECTION TRAY-LF, Model Number: DYNJ19529Q; 9) C-SECTION TRAY-LF, Model Number: DYNJ19529R; 10) C-SECTION TRAY-LF, Model Number: DYNJ19529S; 11) C-SECTION PACK-LF, Model Number: DYNJ21550Q; 12) C-SECTION PACK, Model Number: DYNJ23544F; 13) C-SECTION PACK, Model Number: DYNJ32632F; 14) C-SECTION PACK QVH-LF, Model Number: DYNJ35100I; 15) C-SECTION PACK-LF, Model Number: DYNJ38801F; 16) SCRIPPS SW C-SECTION PACK-LF, Model Number: DYNJ40129C; 17) C-SECTION, Model Number: DYNJ41444F; 18) C-SECTION PACK, Model Number: DYNJ49224I; 19) C SECTION PACK-LF, Model Number: DYNJ52387I; 20) C-SECTION PACK, Model Number: DYNJ56523C; 21) C-SECTION PACK, Model Number: DYNJ59387L; 22) C-SECTION PACK, Model Number: DYNJ64549B; 23) C SECTION PACK, Model Number: DYNJ67347A; 24) C SECTION PACK, Model Number: DYNJ69778A; 25) B-C SECTION BAYSTATE, Model Number: DYNJ900602I; 26) B-C SECTION BAYSTATE, Model Number: DYNJ900602K; 27) PAD C-SECTION, Model Number: DYNJ902134C; 28) RICH C-SECTION, Model Number: DYNJ902182J; 29) C-SECTION PACK, Model Number: DYNJ904329D; 30) C-SECTION, Model Number: DYNJ904697F; 31) KIT C SECTION, Model Number: DYNJ906966G; 32) C SECTION, Model Number: DYNJ907999A; 33) C-SECTION, Model Number: DYNJ909058; 34) C-SECTION PACK, Model Number: DYNJ909507; 35) C SECTION TRAY, Model Number: DYNJT3407; 36) TR-C SECTION PACK-LF, Model Number: PHS583549002B
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number: PAIN0150C
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) HYSTEROSCOPY, Model Number: CDS984086I; 2) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91A; 3) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ0373807J; 4) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ0373807K; 5) MINOR VAGINAL #76-RF, Model Number: DYNJ27434R; 6) VAG HYST PACK, Model Number: DYNJ31344J; 7) VAG HYST PACK, Model Number: DYNJ31344K; 8) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393L; 9) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393M; 10) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393N; 11) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393O; 12) LAVH PACK, Model Number: DYNJ36712F; 13) D&C MINOR LITHOTOMY PACK, Model Number: DYNJ45090I; 14) D C HYSTEROSCOPY PACK-LF, Model Number: DYNJ45801C; 15) VAG HYST PACK (VHGSG)642-LF, Model Number: DYNJ47694I; 16) HYSTEROSCOPY D AND C PACK, Model Number: DYNJ51741B; 17) LAVH - N #668629 -N, Model Number: DYNJ58121C; 18) PK, GYN-MINOR-LITHO, Model Number: DYNJ59037B; 19) HYSTEROSCOPY PACK, Model Number: DYNJ59444G; 20) D AND C PACK, Model Number: DYNJ61282B; 21) VAGINAL HYSTERECTOMY, Model Number: DYNJ67161; 22) PERI GYN, Model Number: DYNJ67708A; 23) ROBOTIC HYSTERECTOMY PACK, Model Number: DYNJ69713B; 24) ROBOTIC HYSTERECTOMY PACK, Model Number: DYNJ69713C; 25) ANORECTAL PACK, Model Number: DYNJ83560; 26) D&C CDS, Model Number: DYNJ902560I; 27) HYSTEROSCOPY, Model Number: DYNJ903327K; 28) ROBOTIC HYST, Model Number: DYNJ903798G; 29) TLH KIT, Model Number: DYNJ908149; 30) LAVH TOTE, Model Number: DYNJ908340; 31) LAVH TOTE, Model Number: DYNJ908340A; 32) LAVH TOTE, Model Number: DYNJ908340B; 33) LAVH TOTE, Model Number: DYNJ908340D; 34) BRECKENRIDGE LAVH, Model Number: DYNJ910537; 35) VAG HYST PACK, Model Number: DYNJT3348; 36) LAVH PART B, Model Number: DYNJV0312G
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) DENTAL PACK, Model Number: DYNDA3003
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/Catalog Number: 380730 Used for minimally invasive surgery.
An error event following the release of a software version for the surgical system that can result in loss of user interface content on an external monitor or tower monitor.
Feb 25, 2026
View details →Olympus High Flow Insufflation Unit. Model Number: UHI. Used for insufflation of the abdominal cavity, and automatic suction and smoke evacuation to facilitate laparoscopic observation and treatment within the abdominal cavity.
Issue with software algorithm which may lead to overpressure events.
Feb 25, 2026
View details →MEDIHONEY CALCIUM ALGINATE WITH ACTIVE LEPTOSPERMUM HONEY. Model Numbers: (1) 31012, (2) 31022, (3) 31045. MEDIHONEY CALCIUM ALGINATE DRESSING WITH ACTIVE LEPTOSPERMUM HONEY supports and aids autolytic debridement and is indicated for the management of moderately to heavily exuding wounds
Potential packaging failures, which could lead to a breach in the sterile barrier.
Feb 25, 2026
View details →Medline Surgical Packs: 1) PACK,UNIVERSAL,ULTRAGARD,5/CS, Model Number: DYNJP1050UG; 2) PACK,SHOULDER,DRAPE,POUCH,10/CSModel Number: DYNJP8401
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) ARTHROSCOPY CDS, Model Number: CDS984609B; 2) ARTHROSCOPY TRAYS, Model Number: DYNJ22501G; 3) ARTHROSCOPY PACK, Model Number: DYNJ32715F; 4) CHS ARTHROSCOPY PACK, Model Number: DYNJ50834C; 5) ARTHROSCOPY TRAY-LF, Model Number: DYNJ51069
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104; 2) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104A; 3) KIT OPHTHALMOLOGY CORNEAL TRAN, Model Number: DYKMBNDL159; 4) KIT OPHTHALMOLOGY CATARACT, Model Number: DYKMBNDL180; 5) KIT OPHTHALMOLOGY EYE MUSCLE, Model Number: DYKMBNDL42; 6) KIT SURG ONC PORT INSERTION, Model Number: DYKMBNDL43B; 7) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84B; 8) KIT OPTHAMOLOGY BASIC, Model Number: DYKMBNDL98; 9) KIT OPTHAMOLOGY BASIC, Model Number: DYKMBNDL98A; 10) KIT OPTHAMOLOGY BASIC, Model Number: DYKMBNDL98B; 11) MINOR EYE PACK, Model Number: DYNJ17219C; 12) PEDS EYE/MUSCLE PACK, Model Number: DYNJ38791M; 13) EYE PACK, Model Number: DYNJ41207C; 14) OPHTHALMIC PACK, Model Number: DYNJ42043I; 15) BREAST RECONSTRUCTION-HOULE, Model Number: DYNJ44852J; 16) CUSTOM EYE PACK HA-LF, Model Number: DYNJ45585C; 17) EYE PLASTIC PACK, Model Number: DYNJ47859B; 18) EYE PLASTIC PACK, Model Number: DYNJ47859C; 19) CATARACT PACK, Model Number: DYNJ53019A; 20) CATARACT PACK, Model Number: DYNJ54869A; 21) VITRECTOMY PACK, Model Number: DYNJ55280B; 22) NDNW-EYE PACK, Model Number: DYNJ61779C; 23) EYE PK, Model Number: DYNJ62306; 24) VITRECTOMY, Model Number: DYNJ64220; 25) OSC BREAST PACK, Model Number: DYNJ65808C; 26) MILLS CATARACT PACK, Model Number: DYNJ67973A; 27) EYE PACK, Model Number: DYNJ68367B; 28) CATARACT PACK, Model Number: DYNJ68875D; 29) CATARACT PACK, Model Number: DYNJ68875F; 30) CATARACT PACK, Model Number: DYNJ68875G; 31) OPHTHO PLASTIC, Model Number: DYNJ902256C; 32) RETINAL, Model Number: DYNJ907495; 33) CATARACT, Model Number: DYNJ909924; 34) COMBO EYE, Model Number: DYNJ909925; 35) VITRECTOMY, Model Number: DYNJ909929
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) PICC LINE TRAY, Model Number: 00-401993O; 2) CODE CART ADULT 4 CENTRAL LINE, Model Number: ACC010268B; 3) CODE CART ADULT 4 CENTRAL LINE, Model Number: ACC010268C; 4) TRAY 6 CENTRAL LINE A, Model Number: ACC010487; 5) PEDS PICC INSERTION TRAY, Model Number: CVI3650; 6) LINE ATTIRE KIT, Model Number: DYKM2170; 7) KIT SURG ONC LAP DIAG/PORT INS, Model Number: DYKMBNDL153C; 8) KIT SURG ONC INSERT CENT.VENOU, Model Number: DYKMBNDL43A; 9) KIT SURG ONC PORT INSERTION, Model Number: DYKMBNDL43C; 10) VEIN PROCEDURE KIT, Model Number: DYNDA2076B; 11) VEIN PROCEDURE KIT, Model Number: DYNDA2076D; 12) VEIN PROCEDURE KIT, Model Number: DYNDA2076F; 13) PLACENTA KIT, Model Number: DYNDA2322B; 14) NICU CENTRAL LINE TRAY, Model Number: DYNDC2407A; 15) NICU CENTRAL LINE TRAY, Model Number: DYNDC2407B; 16) PICC LINE TRAY, Model Number: DYNJ40500A; 17) VENOUS ACCESS PACK, Model Number: DYNJ42694C; 18) AV FISTULA CREATION PACK-LF, Model Number: DYNJ53427B; 19) PICC LINE, Model Number: DYNJ55546D; 20) PICC TRAY, Model Number: DYNJ63199A; 21) PICC KIT 4F SL PL MAX BARRIER, Model Number: DYNJ70141MB; 22) PICC KIT 5F DL PL MB, Model Number: DYNJ70152MB; 23) VEIN PACK, Model Number: DYNJ84239; 24) PICC INSERTION TRAY W/O CATH, Model Number: PICCNC0004
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) TOTE ABRAMS KNEE ARTHROSCOPY, Model Number: DYKM1801B; 2) TOTE SEN OSC HAND REPAIR, Model Number: DYKM1776; 3) TOTE MOMENI SB CAPSULOTOMY REC, Model Number: DYKM1765A; 4) TOTE YAO OSC HAND REPAIR, Model Number: DYKM1796A; 5) TOTE SB NGUYEN CAPSULOTOMY, Model Number: DYKM1756A; 6) TOTE FOX OSC HAND REPAIR, Model Number: DYKM1783A; 7) TOTE KAMAL OSC HAND REPAIR, Model Number: DYKM1730A; 8) TOTE SB NGUYEN LIPOSUCTION, Model Number: DYKM1927A; 9) TOTE YAO OSC WRIST ARTHROSCOPY, Model Number: DYKM1807; 10) TOTE CURTIN OSC HAND REPAIR, Model Number: DYKM1781A; 11) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977B; 12) TOTE CHANGE OSC HAND REPAIR, Model Number: DYKM1751A; 13) TOTE CARROLL BLOCK BLOOD PATCH, Model Number: DYKM1864; 14) TOTE SB STONE JWIRE LUMPECTOMY/SEN NODE, Model Number: DYKM1925; 15) TOTE SAFRAN OSC HIP ARTHRSCPY, Model Number: DYKM1767B; 16) CVC PACK, Model Number: CVI4500A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.
Feb 25, 2026
View details →MOLLI 2 System System includes: MM1000 (Pack of 10) - MOLLI Introducer (8cm, 14G) and MOLLI Marker UDI (GTIN): 00850024195028 MS-00009 (Pack of 10) - MOLLI Introducer (12cm, 14G) and MOLLI Marker UDI (GTIN): 00850024195134 MP2001 - MOLLI 2 Wand UDI (GTIN): 00850024195141 MP2015A - MOLLI OncoPen UDI (GTIN): 00850024195219 MT2001 - MOLLI 2 Tablet UDI (GTIN): 00850024195158 The MOLLI Marker is intended to be placed percutaneously in soft tissue to temporarily mark a surgical site intended for surgical removal. the MOLLI Marker is located and surgically removed with the target tissue.
Due to a need to update the product labeling (Instructions for Use IFU) as a result of the magnetized surgical tools dislodging markers using during target tissue removal procedures.
Feb 25, 2026
View details →Medline Convenience Kits: 1) ENDO KIT W/SYRINGE, Model Number: DYKE1743; 2) BGMC ANGIOGRAPHY PACK-LF, Model Number: DYNJ0800934D; 3) GENERAL ENDO PACK-LF, Model Number: DYNJ0842873J; 4) OR ANGIO PACK-LF, Model Number: DYNJ43415C; 5) ANGIO PACK II, Model Number: DYNJ44066A; 6) SOUTH CATH LAB PACK, Model Number: DYNJ61978A; 7) BIOPSY ANGIO TRAY ALEX, Model Number: DYNJ63542B; 8) SURGICAL TECH KIT, Model Number: DYNJ907452; 9) ANGIOGRAPHY PACK, Model Number: DYNJC2371G; 10) SPECTRUM HEALTH CATH PACK-LF, Model Number: PHS853965G
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138; 4) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142; 5) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142B; 6) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145; 7) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145B; 8) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5; 9) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5B; 10) NEURO PACK-LF, Model Number: DYNJ0305809O; 11) NEURO PACK-LF, Model Number: DYNJ0305809P; 12) NEURO PACK-LF, Model Number: DYNJ0408776Y; 13) NEURO PACK-LF, Model Number: DYNJ04596K; 14) BASIC NEURO PACK SMMC-LF, Model Number: DYNJ35243L; 15) NEURO PACK, Model Number: DYNJ49098I; 16) NEURO PACK, Model Number: DYNJ53089B; 17) NEURO LAMINECTOMY-LF, Model Number: DYNJ53799; 18) NEURO MINOR PACK-UNIV, Model Number: DYNJ58704A; 19) NEURO PACK - IFCAP 230316, Model Number: DYNJ60100C; 20) NEURO PACK, Model Number: DYNJ60792B; 21) NEURO PACK, Model Number: DYNJ60792C; 22) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157C; 23) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157D; 24) CHRISTUS CHILDRENS NEURO PK, Model Number: DYNJ61168D; 25) NEURO PACK, Model Number: DYNJ61740D; 26) NEURO ST DAVIDS MEDICAL CTR, Model Number: DYNJ61998; 27) NEURO GOODIE PACK - DUNN, Model Number: DYNJ64570A; 28) DISCECTOMY PACK, Model Number: DYNJ67835; 29) WWD NEURO PACK, Model Number: DYNJ80849B; 30) WWD NEURO PACK, Model Number: DYNJ80849J; 31) NEURO DIAGNOSTIC, Model Number: DYNJ84653A; 32) NEURO, Model Number: DYNJ900904J; 33) NEURO, Model Number: DYNJ900904K; 34) NEURO, Model Number: DYNJ900904L; 35) NEURO, Model Number: DYNJ900904M; 36) NEURO, Model Number: DYNJ902585; 37) NEURO, Model Number: DYNJ902585A; 38) NEURO - IFCAP 230316, Model Number: DYNJ908393; 39) NEURO - IFCAP 230316, Model Number: DYNJ908393A; 40) NEURO KIT, Model Number: DYNJ909573; 41) NEURO, Model Number: DYNJ910384; 42) NEURO, Model Number: DYNJ910384A; 43) NEURO PACK-LF, Model Number: PHS390015M
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) BAPTIST FLOYD ENDO GI KIT, Model Number: DYKE1455B; 2) GI LAB OTHER ENDO KIT, Model Number: DYKE1721
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL31; 4) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38; 5) L & D PACK-LF, Model Number: DYNJ0456816O; 6) L & D PACK-LF, Model Number: DYNJ0456816P; 7) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ04692S; 8) VAGINAL DELIVERY PACK, Model Number: DYNJ16621B; 9) VAGINAL DELIVERY PACK, Model Number: DYNJ16621C; 10) DELIVERY PACK-LF, Model Number: DYNJ20546F; 11) VAGINAL DELIVERY PK, Model Number: DYNJ20979I; 12) VAGINAL DELIVERY PK, Model Number: DYNJ20979J; 13) VAGINAL DELIVERY PACK, Model Number: DYNJ24108D; 14) OB PACK-LF, Model Number: DYNJ26121I; 15) OB DELIVERY PACK, Model Number: DYNJ31024J; 16) OB DELIVERY PACK-LF, Model Number: DYNJ32139C; 17) DB VAG HYSTERECTOMY-LF, Model Number: DYNJ33590F; 18) MAJOR VAGINAL PACK, Model Number: DYNJ35488F; 19) ENS. ACCOUCHEMENT VAGINAL-LF, Model Number: DYNJ37331C; 20) VAG DELIVERY PACK, Model Number: DYNJ38481G; 21) VAG DELIVERY PACK, Model Number: DYNJ38481I; 22) VAG DELIVERY PACK, Model Number: DYNJ38481K; 23) OB PACK 2, Model Number: DYNJ39000D; 24) GYNECOLOGY VAGINAL KIT, Model Number: DYNJ40203A; 25) LABOR DELIVERY POUCH SCREEN, Model Number: DYNJ40918A; 26) L & D PACK WITHOUT INSTRUMENTS, Model Number: DYNJ44720B; 27) VAGINAL DELIVERY PACK, Model Number: DYNJ51377G; 28) FH VAG DELIVERY PK-LF, Model Number: DYNJ51794B; 29) VAGINAL DELIVERY PACK, Model Number: DYNJ55363A; 30) VAGINAL DELIVERY PACK, Model Number: DYNJ55363B; 31) MINOR VAGINAL PACK, Model Number: DYNJ61269B; 32) OB VAG DELIVERY PACK, Model Number: DYNJ61574; 33) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ61650; 34) OB-VAG PACK 9128190, Model Number: DYNJ62341; 35) HP VAGINAL DELIVERY PACK-LF, Model Number: DYNJ62402A; 36) L & D PACK, Model Number: DYNJ63216A; 37) L&D PACK, Model Number: DYNJ63757A; 38) GYN PACK DYNJ55833A, Model Number: DYNJ65145; 39) MINOR VAG PACK, Model Number: DYNJ68663; 40) UTHET TYLER OB PACK, Model Number: DYNJ68673A; 41) GYN VAGINAL PACK, Model Number: DYNJ68681; 42) VAGINAL DELIVERY PACK, Model Number: DYNJ69544A; 43) VAGINAL PACK, Model Number: DYNJ69774A; 44) LABOR & DELIVERY PACK-LF, Model Number: DYNJ84393; 45) DELIVERY PACK, Model Number: DYNJ87469A; 46) VAGINAL DELIVERY, Model Number: DYNJ901387A; 47) FLOYD DELIVERY, Model Number: DYNJ906937D; 48) OB VAGINAL MEDSTAR PACK-LF, Model Number: DYNJT3445; 49) OB PACK, Model Number: DYNJT5186; 50) VAGINAL DELIVERY KIT, Model Number: MNS13550A; 51) VAGINAL DELIVERY TRAY, Model Number: MNS7460A; 52) LABOR AND DELIVERY KIT, Model Number: MNS8950A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Olympus Thunderbeat 5 mm, 10 cm, Inline Grip
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 3) KIT PLASTIC INSERT TISSUE EXPA, Model Number: DYKMBNDL158; 4) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170; 5) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170A; 6) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170B; 7) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182; 8) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182A; 9) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182B; 10) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183; 11) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183A; 12) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183B; 13) KIT STC FACIAL FRACTURE, Model Number: DYKMBNDL34A; 14) KIT SURG ONC MASTECTOMY A, Model Number: DYKMBNDL39; 15) KIT SURG ONC MASTECTOMY A, Model Number: DYKMBNDL39A; 16) KIT SURG ONC MASTECTOMY, Model Number: DYKMBNDL39C; 17) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL46; 18) KIT STC PLASTIC FREE FLAP, Model Number: DYKMBNDL54; 19) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69; 20) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69A; 21) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69B; 22) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78; 23) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78A; 24) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78B; 25) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84; 26) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84C; 27) MJR PLASTICS PK MV SURG CT-LF, Model Number: DYNJ0785310G; 28) ASC ENT PLASTIC PACK-LF, Model Number: DYNJ0854650J; 29) BREAST PACK ASC-LF, Model Number: DYNJ36469I; 30) ENT TYMPANOPLASTY PACK-LF, Model Number: DYNJ38988J; 31) DR WILDE BRSTMASS/PARTMAST-XM, Model Number: DYNJ39831F; 32) BREAST RECONSTRUCTION-HOULE, Model Number: DYNJ44852K; 33) BREAST RECONSTRUCTION-HOULE, Model Number: DYNJ44852M; 34) BREAST PACK, Model Number: DYNJ50957G; 35) PLASTICS BREAST PACK, Model Number: DYNJ55358G; 36) BHN IMPLANT PACK, Model Number: DYNJ56737A; 37) BREAST PACK, Model Number: DYNJ60210A; 38) PLASTICS PACK 2, Model Number: DYNJ64927F; 39) SPECIALTY FACIAL PLASTICS PACK, Model Number: DYNJ68844A; 40) BREAST IMPLANT PACK, Model Number: DYNJ69351; 41) BREAST/TUMMY PACK, Model Number: DYNJ83788; 42) HMSL PLASTICS, Model Number: DYNJ83884B; 43) CHEST/BREAST-LF, Model Number: DYNJ83950A; 44) CHEST/BREAST-LF, Model Number: DYNJ83950B; 45) FREE FLAP A DIRTY, Model Number: DYNJ86165; 46) MAJOR SURGERY PACK, Model Number: DYNJ86816; 47) BREAST REDUCTION PACK, Model Number: DYNJ87444; 48) OMF FREE FLAP, Model Number: DYNJ902005I; 49) BREAST CDS, Model Number: DYNJ902558G; 50) TRUNCAL CONTOURING, Model Number: DYNJ903555D; 51) TRUNCAL CONTOURING, Model Number: DYNJ903555F; 52) NEW TISSUE RETRIEVAL MH, Model Number: DYNJ903849C; 53) LARGE PROCEDURE-LF, Model Number: DYNJ907847A; 54) (34) STC FACIAL FRACTURE, Model Number: DYNJ907874A; 55) FACELIFT, Model Number: DYNJ910462; 56) MAJOR PLASTIC BREAST PACK-LF, Model Number: DYNJT3283; 57) MAJOR BREAST PACK-LF, Model Number: DYNJT3417; 58) PLASTICS PACK-LF, Model Number: PHS806973G
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) BURN PACK, Model Number: DYNJ15668R; 2) BURN PACK-LF, Model Number: DYNJ42829B; 3) BURN CARE PACK-LF, Model Number: DYNJT3437
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Surgical Gowns: 1) GOWN,NONRNF,L,30/CS, Model Number: DYNJP2001; 2) GOWN,SIRUS,NONRNF,SETINSLV,L,20/CS, Model Number: DYNJP2001S; 3) GOWN,ECLIPSE,NONRNF,XL,ST,30/CS, Model Number: DYNJP2002; 4) GOWN,SIRUS,NONRNF,SETINSLV,XL,20/CS, Model Number: DYNJP2002S; 5) GOWN,SIRUS,NONRNF,XLN/XL,20/CS, Model Number: DYNJP2002SL; 6) GOWN,NONRNF,2XL,18/CS, Model Number: DYNJP2003; 7) GOWN,SIRUS,NONRNF,SETINSLV,2XL,18/CS, Model Number: DYNJP2003S; 8) GOWN,SIRUS,NONRNF,XLN/2XL,18/CS, Model Number: DYNJP2003SL; 9) GOWN,NONRNF,3XL,18/CS, Model Number: DYNJP2004; 10) GOWN,SIRUS,NONRNF,3XL,18/CS, Model Number: DYNJP2004S; 11) GOWN,NONRNF,SM,30/CS, Model Number: DYNJP2005; 12) GOWN,SIRUS,NONRNF,SM,30/CS, Model Number: DYNJP2005S; 13) GOWN,NONRNF,4XL,18/CS, Model Number: DYNJP2009; 14) GOWN,SIRUS,NONRNF,4XL,18/CS, Model Number: DYNJP2009S; 15) GOWN,ECLIPSE,FABRNF,L,30/CS, Model Number: DYNJP2101; 16) GOWN,SIRUS,FABRNF,L,20/CS, Model Number: DYNJP2101S; 17) GOWN,ECLIPSE,FABRNF,XL,30/CS, Model Number: DYNJP2102; 18) GOWN,SIRUS,FABRNF,XL,20/CS, Model Number: DYNJP2102S; 19) GOWN,ECLIPSE,FABRNF,2XL,18/CS, Model Number: DYNJP2103; 20) GOWN,SIRUS,FABRNF,2XL,18/CS, Model Number: DYNJP2103S; 21) GOWN,ECLIPSE,POLYRNF,L,30/CS, Model Number: DYNJP2201; 22) GOWN,SIRUS,POLYRNF,SETINSLV,L,20/CS, Model Number: DYNJP2201S; 23) GOWN,ECLIPSE,POLYRNF,XL,30/CS, Model Number: DYNJP2202; 24) GOWN,SIRUS,POLYRNF,SETINSLV,XL,20/CS, Model Number: DYNJP2202S; 25) GOWN,ECLIPSE,POLYRNF,2XL,18/CS, Model Number: DYNJP2203; 26) GOWN,SIRUS,POLYRNF,SETINSLV,2XL,18/CS, Model Number: DYNJP2203S; 27) GOWN,ECLIPSE,POLYRNF,XLN/L,28/CS, Model Number: DYNJP2204; 28) GOWN,SIRUS,POLYRNF,XLN/LG,20/CS, Model Number: DYNJP2204S; 29) GOWN,ECLIPSE,POLYRNF,XLN/XL,28/CS, Model Number: DYNJP2205; 30) GOWN,SIRUS,POLYRNF,XLN/XL,20/CS, Model Number: DYNJP2205S; 31) GOWN,ECLIPSE,POLYRNF,XLN/2XL,18/CS, Model Number: DYNJP2206; 32) GOWN,SIRUS,POLYRNF,XLN/2XL,18/CS, Model Number: DYNJP2206S; 33) DBQ-GOWN,PREVENTION PLUS,LN/L,ST,24/CS, Model Number: DYNJP2301P; 34) DBQ-GOWN,PREVENTION PLUS,XLN/XL,ST,24/CS, Model Number: DYNJP2302P; 35) DBD-GOWN,PREVENTION PLUS,XLN/2XL,ST,22/C, Model Number: DYNJP2303P; 36) DBD-GOWN,PREVENTION PLUS,L,ST,24/CS, Model Number: DYNJP2306P; 37) DBQ-GOWN,PREVENTION PLUS,XL,ST,24/CS, Model Number: DYNJP2307P; 38) DBQ-GOWN,PREVENTION PLUS,2XL,ST,22/CS, Model Number: DYNJP2308P; 39) GOWN,ORBIS,LVL 3,LRG/XLONG,ST,24/CS, Model Number: DYNJP2361P; 40) GOWN,ORBIS,LVL 3,XLONG/XLARGE,ST,24/CS, Model Number: DYNJP2362P; 41) GOWN,ORBIS,LVL 3,XLNG/XXLARGE,ST,22/CS, Model Number: DYNJP2363P; 42) GOWN,ORBIS,LVL 3,LARGE,ST,24/CS, Model Number: DYNJP2366P; 43) GOWN,ORBIS,LVL 3,XLARGE,ST,24/CS, Model Number: DYNJP2367P; 44) GOWN,ORBIS,LVL 3,XXLARGE,ST,22/CS, Model Number: DYNJP2368P; 45) GOWN,ORBIS,LVL 3,4XL,ST,20/CS, Model Number: DYNJP2369P; 46) GOWN,SIRUS,NONRNF,RAGLAN,L,ST,32/CS, Model Number: DYNJP2401; 47) GOWN,SIRUS,NONRNF,RAGLAN,XL,ST,30/CS, Model Number: DYNJP2402; 48) GOWN,SIRUS,NONRNF,RAGLAN,2XL,ST,28/CS, Model Number: DYNJP2403; 49) GOWN,SIRUS,FABRNF,RAGLAN,L,ST,30/CS, Model Number: DYNJP2501; 50) GOWN,SIRUS,FABRNF,RAGLAN,XL,ST,28/CS, Model Number: DYNJP2502; 51) GOWN,SIRUS,FABRNF,RAGLAN,2XL,ST,28/CS, Model Number: DYNJP2503; 52
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Olympus Thunderbeat, 5 mm, 45 cm, Pistol Grip
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Olympus Thunderbeat 5 mm, 35 cm, Front-Actuated Grip
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Medline Convenience Kits: 1) ROBOTIC GYN CDS-1, Model Number: CDS982893I; 2) GYN LAPAROSCOPY CDS, Model Number: CDS983300J; 3) NHP ROBOTIC GYNE CDS, Model Number: CDS984243G; 4) ROBOTIC-LF, Model Number: CDS984543D; 5) ROBOTIC-LF, Model Number: CDS984543F; 6) ROBOTIC-LF, Model Number: CDS984543G; 7) ROBOTIC-LF, Model Number: CDS984543I; 8) GYN CDS, Model Number: CDS984860F; 9) GYN LAP PELVI, Model Number: CDS985423G; 10) GYN KIT, Model Number: DYKM1657A; 11) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL100; 12) KIT ROBOTICS GYN, Model Number: DYKMBNDL109; 13) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116D; 14) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116F; 15) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119; 16) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119A; 17) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168; 18) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168A; 19) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184D; 20) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91; 21) KIT GYN ONCOLOGY, Model Number: DYKMBNDL92; 22) KIT GYN ONCOLOGY, Model Number: DYKMBNDL92A; 23) KIT GYN ABDOMINAL, Model Number: DYKMBNDL92C; 24) KIT GYN LAP, Model Number: DYKMBNDL94; 25) KIT GYN LAP, Model Number: DYKMBNDL94A; 26) KIT GYN LAP, Model Number: DYKMBNDL94C; 27) KIT GYN LAP, Model Number: DYKMBNDL94D; 28) GYN LAPAROSCOPY PACK-LF, Model Number: DYNJ0281948N; 29) PROSTATE ROBOTIC PACK, Model Number: DYNJ0347372F; 30) PROSTATE ROBOTIC PACK, Model Number: DYNJ0347372G; 31) PROSTATE ROBOTIC PACK, Model Number: DYNJ0347372I; 32) HEALTH ALL GYN LAP PACK-LF, Model Number: DYNJ28252G; 33) GYN OPERATIVE LAP PACK-LF, Model Number: DYNJ30005G; 34) OB PACK, Model Number: DYNJ32706B; 35) ROBOTIC UROLOGY-RF, Model Number: DYNJ38844N; 36) ENSEMBLE GYNE LAPAROTOMIE-LF, Model Number: DYNJ47550C; 37) GYN LAP PACK, Model Number: DYNJ47714A; 38) TASC GYN LAP, Model Number: DYNJ49191B; 39) ROBOTIC PACK, Model Number: DYNJ49705K; 40) GYN LAPAROSCOPY PACK, Model Number: DYNJ50348F; 41) NWH GYN LAPAROSCOPY, Model Number: DYNJ50439F; 42) FH LAPAROSCOPIC GYN PK-LF, Model Number: DYNJ51795F; 43) GYN LAPAROSCOPY PACK, Model Number: DYNJ52569B; 44) LAVH PACK, Model Number: DYNJ55862; 45) PK, GEN-LAPAROSCOPY, Model Number: DYNJ57607A; 46) GYN LAPAROSCOPY MRNGSD, Model Number: DYNJ58321A; 47) GYN LAPAROSCOPY PACK, Model Number: DYNJ63079B; 48) GYN LAPAROSCOPY PACK, Model Number: DYNJ63079C; 49) GYN LAPAROSCOPY PACK, Model Number: DYNJ63079D; 50) GYN LAPAROSCOPY PACK, Model Number: DYNJ63079F; 51) GYN DAVINCI PACK, Model Number: DYNJ64266; 52) DAVINCI PACK, Model Number: DYNJ66298D; 53) GYN LAPAROSCOPY, Model Number: DYNJ67269; 54) GENERAL GYN PACK, Model Number: DYNJ68710; 55) LAP GYN PACK, Model Number: DYNJ69419A; 56) LAP GYN PACK, Model Number: DYNJ69419D; 57) GYN LAPAROSCOPY, Model Number: DYNJ81492A; 58) PK, ROBOTIC GYN & GU-ROSE MEDI, Model Number: DYNJ81761A; 59) GYN PACK, Model Number: DYNJ83475A; 60) ROBOTIC XI PACK, Model Number: DYNJ84783; 61) GYN PACK, Model Number: DYNJ86195A; 62) GYN LAPAROSCOPY, Model Number: DYNJ900262A; 63) D&C HYSTEROSCOPY, Model Number: DYNJ901133A; 64) GYN LAPAROSCOPY, Model Number: DYNJ901559J; 65) GYN ENDOSCOPY-LF, Model Number: DYNJ901812J; 66) GYN LITHOTOMY-LF, Model Number: DYNJ902711M; 67) GENERAL ROBOTIC, Model Number: DYNJ905066B; 68) TRINITY C-SEC W/LINEN CDS, Model Number: DYNJ905269G; 69) KIT GYN LAPAROSCOPY RFD, Model Number: DYNJ906322D; 70) GYN LAP LH, Model Number: DYNJ906452C; 71) GYN LAP LH, Model Number: DYNJ906452D; 72) GYN LAP LH, Model Number: DYNJ906452F; 73) FLOYD LITH GYN, Model Number: DYNJ906926B; 74) GYN LAPAROSCOPY-MRMC, Model Number: DYNJ907753F; 75) GYN LAPAROSCOPY-SFMC, Model Number: DYNJ907771C; 76) LAP/GYN ROBOTICS, Model Number: DYNJ908887B; 77) LAP/GYN ROBOTICS, Model Number: DYNJ908887C;
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Description: Infusion Pump, one unit per carton Component: N/A
Pumps were released without full testing being performed, including occlusion alarm testing.
Feb 25, 2026
View details →Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035A; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A;
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) WMC I D PACK-LF, Model Number: DYNJ50806L
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) PPE KIT, Model Number: DYK1011945P
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK CDS-LF, Model Number: CDS983782C; 4) BAPTIST FLOYD BRONCHOSCOPY, Model Number: DYKE1456D; 5) BRONCHOSCOPY SETUP KIT, Model Number: DYKE1955; 6) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156; 7) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156A; 8) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156B; 9) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174; 10) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174A; 11) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174B; 12) KIT PEDS HEAD AND NECK, Model Number: DYKMBNDL196; 13) KIT OMF LEFORTE, Model Number: DYKMBNDL1A; 14) KIT OMF LEFORTE, Model Number: DYKMBNDL1B; 15) KIT OMF LEFORTE, Model Number: DYKMBNDL1C; 16) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32; 17) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32A; 18) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32B; 19) KIT GEN SURG THYROIDECTOMY, Model Number: DYKMBNDL66A; 20) KIT GEN SURG THYROIDECTOMY, Model Number: DYKMBNDL66B; 21) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88A; 22) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88B; 23) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88C; 24) KIT GEN SURG LAP VENTRA, Model Number: DYKMBNDL96C; 25) MINOR ENT PACK, Model Number: DYNJ01831I; 26) ENT BASIC PACK-LF, Model Number: DYNJ0387711F; 27) D AND C PACK-LF, Model Number: DYNJ0855090P; 28) CSS SINUS PACK, Model Number: DYNJ17005I; 29) CSS HEAD & NECK PACK, Model Number: DYNJ26780I; 30) NASALPLASTY #61-RF, Model Number: DYNJ27320V; 31) MASTOID PACK-WINTER PARK-LF, Model Number: DYNJ32060C; 32) T&A PACK, Model Number: DYNJ32809I; 33) HEAD AND NECK PACK, Model Number: DYNJ41169; 34) SIMPLE NECK DISSECTION, Model Number: DYNJ47481C; 35) ENT PACK, Model Number: DYNJ47882F; 36) ENT PACK, Model Number: DYNJ48401C; 37) ENT PACK, Model Number: DYNJ51934A; 38) PK,ORTHO-EXTREMITY-LOSROBLES, Model Number: DYNJ52916A; 39) HEAD & NECK PACK, Model Number: DYNJ54964A; 40) HEAD & NECK PACK, Model Number: DYNJ54964B; 41) ENT - MINOR ENT PACK-LF, Model Number: DYNJ56840B; 42) GLENNON HEAD AND NECK PACK-LF, Model Number: DYNJ57200C; 43) ENT PACK, Model Number: DYNJ57688F; 44) RF T AND A PACK, Model Number: DYNJ61393A; 45) T & A PACK, Model Number: DYNJ61935D; 46) ENT PACK, Model Number: DYNJ62093B; 47) ENT PACK, Model Number: DYNJ64024B; 48) OSCLJ HEAD AND NECK PACK, Model Number: DYNJ66509A; 49) HEAD AND NECK PACK, Model Number: DYNJ67208; 50) HEAD & NECK PACK, Model Number: DYNJ67328B; 51) PEDS HEAD AND NECK PACK, Model Number: DYNJ69746B; 52) NASAL PACK, Model Number: DYNJ81280A; 53) ENT-PLASTICS PACK, Model Number: DYNJ84258A; 54) ENT-PLASTICS PACK, Model Number: DYNJ84258B; 55) HEAD & NECK PACK, Model Number: DYNJ85078A; 56) ENT PACK, Model Number: DYNJ86194A; 57) EENT PACK, Model Number: DYNJ86235; 58) ENT PACK, Model Number: DYNJ87378; 59) ENT PACK, Model Number: DYNJ87378A; 60) ENT PACK, Model Number: DYNJ87378D; 61) ENT SPLIT PACK, Model Number: DYNJ87451; 62) ENT SINUS PACK, Model Number: DYNJ88498; 63) NASAL, Model Number: DYNJ902993F; 64) ENDO SINUS-MIDTOWN, Model Number: DYNJ905866; 65) T A, Model Number: DYNJ906292B; 66) THYROID, Model Number: DYNJ906841A; 67) THYROID - SMH AMB, Model Number: DYNJ907739B; 68) ENT, Model Number: DYNJ911623; 69) HEAD AND NECK PACK, Model Number: DYNJT5739; 70) RO MEDIASTINAL PACK-LF, Model Number: PHS393087007C
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Olympus Thunderbeat, 5 mm, 20 cm, Inline Grip
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number: DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number: DYNJ905567A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Olympus High Flow Insufflation Unit. Model/ Number: UHI-2. Used for insufflation of the abdominal cavity, and automatic suction and smoke evacuation to facilitate laparoscopic observation and treatment within the abdominal cavity.
Issue with software algorithm which may lead to overpressure events.
Feb 25, 2026
View details →Medline Convenience Kits: 1) KIT STC TRACHEOSTOMY, Model Number: DYKMBNDL55; 2) KIT STC TRACHEOSTOMY, Model Number: DYKMBNDL55A; 3) TRACH TOTE, Model Number: DYNJ85691
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROTOMY CDS, Model Number: CDS860015V; 3) MAJOR LAPAROTOMY CDS, Model Number: CDS860015W; 4) MAJOR LAPAROTOMY CDS, Model Number: CDS860015X; 5) MINOR LAPAROTOMY CDS, Model Number: CDS860016P; 6) MINOR LAPAROTOMY CDS, Model Number: CDS860016Q; 7) MINOR LAPAROTOMY CDS, Model Number: CDS860016R; 8) MINOR LAPAROTOMY CDS, Model Number: CDS860016S; 9) LAP GASTRIC BYPASS CDS, Model Number: CDS860146M; 10) NEW LONDON MINOR CDS, Model Number: CDS982518M; 11) NEW LONDON MINOR CDS, Model Number: CDS982518N; 12) NEW LONDON MINOR CDS, Model Number: CDS982518O; 13) NEW LONDON MINOR CDS, Model Number: CDS982518P; 14) GENERAL LAPAROSCOPY CDS, Model Number: CDS983311J; 15) LAVH CDS, Model Number: CDS983411G; 16) NHP ROBOTIC GENERAL CDS, Model Number: CDS984244J; 17) ROBOTIC-LF, Model Number: CDS984543C; 18) CYSTO CDS, Model Number: CDS984592G; 19) CYSTO CDS, Model Number: CDS984592I; 20) ROBOTIC, Model Number: CDS985270F; 21) GENERAL LAPAROSCOPY CDS, Model Number: CDS985422I; 22) ROBOTIC CDS, Model Number: CDS985429L; 23) LAP CHOLE CDS, Model Number: CDS985558N; 24) LAP CHOLE CDS, Model Number: CDS985558O; 25) ENDO KIT, Model Number: DYKE1580A; 26) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105; 27) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105A; 28) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105B; 29) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116; 30) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116B; 31) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117A; 32) KIT SURG ONC DIAGNOSTIC LAPARO, Model Number: DYKMBNDL12; 33) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121; 34) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121A; 35) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121B; 36) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121C; 37) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121F; 38) KIT L&D CYSTO SETUP, Model Number: DYKMBNDL135A; 39) KIT THORACIC LAPAROSCOPIC HIAT, Model Number: DYKMBNDL136; 40) KIT THORACIC LAPAROSCOPIC HIAT, Model Number: DYKMBNDL136A; 41) KIT THORACIC STAGING LAPAROTOM, Model Number: DYKMBNDL15; 42) KIT SURG ONC LAP DIAG/CV CATH, Model Number: DYKMBNDL153A; 43) KIT SURG ONC LAP DIAG/PORT INS, Model Number: DYKMBNDL153B; 44) KIT LAP DONOR NEPHRECTOMY, Model Number: DYKMBNDL173; 45) KIT LAP DONOR NEPHRECTOMY, Model Number: DYKMBNDL173A; 46) KIT THORACIC HIATAL HERNIA TRA, Model Number: DYKMBNDL175; 47) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176; 48) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176A; 49) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176B; 50) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176C; 51) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184; 52) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184A; 53) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188A; 54) KIT UROLOGY LAP SIDDIQUI, Model Number: DYKMBNDL193; 55) KIT UROLOGY LAP SIDDIQUI, Model Number: DYKMBNDL193A; 56) KIT UROLOGY BUCCAL GRAFT, Model Number: DYKMBNDL206; 57) KIT STC PLASTIC FREE FLAP, Model Number: DYKMBNDL54A; 58) KIT UROLOGY CYSTOSCOPY SUPPLEM, Model Number: DYKMBNDL60; 59) KIT UROLOGY CYSTOSCOPY SUPPLEM, Model Number: DYKMBNDL60A; 60) KIT UROLOGY MINOR, Model Number: DYKMBNDL61A; 61) KIT UROLOGY MINOR, Model Number: DYKMBNDL61B; 62) KIT GEN SURG LAP COLECTOMY, Model Number: DYKMBNDL73; 63) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73A; 64) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73B; 65) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73C; 66) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73D; 67) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74; 68) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74A; 69) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74B; 70) KIT GEN SURG LAP GASTRIC BYPAS,
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) KIT GENERAL CLOSURE, Model Number: DYKMBNDL200A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Olympus High Flow Insufflation Unit. Model Number: UHI-3. Used for insufflation of the abdominal cavity, and automatic suction and smoke evacuation to facilitate laparoscopic observation and treatment within the abdominal cavity.
Issue with software algorithm which may lead to overpressure events.
Feb 25, 2026
View details →Medline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 2) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431K; 3) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431O; 4) KIT ROBOTICS GYN, Model Number: DYKMBNDL109A; 5) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116A; 6) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116G; 7) ROBOTIC PACK, Model Number: DYNJ04135B; 8) MINOR LITHOTOMY PACK, Model Number: DYNJ04382J; 9) NBMC D & C PACK-LF, Model Number: DYNJ0665843C; 10) D&C PACK-LF, Model Number: DYNJ22567G; 11) D&C PACK-LF, Model Number: DYNJ22567J; 12) D AND C, Model Number: DYNJ41272B; 13) D AND C, Model Number: DYNJ41272C; 14) OB PACK, Model Number: DYNJ42884; 15) D&C/GYN PACK, Model Number: DYNJ47713A; 16) PERI GYN PACK, Model Number: DYNJ55377D; 17) ANTERIOR CERVICAL-LF, Model Number: DYNJ58344; 18) D&C PACK, Model Number: DYNJ67214D; 19) D&C PACK, Model Number: DYNJ67214F; 20) D&C PACK, Model Number: DYNJ67214G; 21) D&C PACK, Model Number: DYNJ67214I; 22) LITHOTOMY PACK, Model Number: DYNJ80765C; 23) GYN PACK, Model Number: DYNJ83475; 24) D AND C PACK, Model Number: DYNJ87468; 25) D&C-ASC, Model Number: DYNJ900465C; 26) D&C-ASC, Model Number: DYNJ900465D; 27) D&C-ASC, Model Number: DYNJ900465F; 28) GYN LITHOTOMY-LF, Model Number: DYNJ902711L; 29) GYN LITHOTOMY-LF, Model Number: DYNJ902711N; 30) D&C, Model Number: DYNJ906947C; 31) PACK PERI GYN DSMC, Model Number: DYNJT6424
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: TESLA-B2535-450 Model Number: G60785 UDI-DI code: 00827002607859 These devices are used for cannulation of the ductal system and for sphincterotomy. If preloaded, the device also aids in bridging difficult strictures during ERCP. It is indicated for sphincterotome-aided, wire-guided selective cannulation of the biliary ducts.
Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.
Feb 25, 2026
View details →Private Label CVS. Model Number: CVS405406. Helps maintain a moist wound environment, which has shown to be conducive to wound healing. Intended for minor scalds & burns, minor cuts, lacerations & minor abrasions.
Potential packaging failures, which could lead to a breach in the sterile barrier.
Feb 25, 2026
View details →MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software
Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.
Feb 25, 2026
View details →Medline Convenience Kits: 1) ADMIT KIT, Model Number: DYKA1343A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Olympus Thunderbeat 5 mm, 20 cm, Front-Actuated Grip Type S
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model Number: DYNDTB0529; 2) IV ADMIN SET, 10DRP, 1NF, 92, Model Number: DYNDTB1029; 3) DBD-IVEXTSET2.4ML2 NEEDLE-FREEVLVE,35, Model Number: DYNDTB1516; 4) DBD-IV ADMIN SET, 15DRP, 1 INJECT PORT,, Model Number: DYNDTB1530; 5) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 6, Model Number: DYNDTB1539; 6) DBD-IV ADMIN SET, 15DRP, 3 NF, 109, Model Number: DYNDTB1545; 7) IV ADMIN SET, 20DRP, 3 NF, 109, Model Number: DYNDTB2045; 8) DBD-IVADMINSET60DRP2NEEDLE-FREEVLVE96, Model Number: DYNDTB0512; 9) DBD-IVADMINSET,60DRP3 NEEDLE-FREE VLVE11, Model Number: DYNDTB0555; 10) IV ADMIN SET, 10DRP, NF,INJ, 96, Model Number: DYNDTB1022; 11) DBD-IV ADMIN SET, 15DRP, 2NF, 96, Model Number: DYNDTB1512; 12) DBD-IV ADMIN SET, 15DRP, 1NF, 92, Model Number: DYNDTB1529; 13) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 9, Model Number: DYNDTB1537; 14) DBD-IV ADMIN SET, 15DRP, SECOND, 40, Model Number: DYNDTB1540; 15) IV ADMIN SET, 20DRP, 1NF, 92, Model Number: DYNDTB2029; 16) DBD-IV EXT SET, 6.5, 1.1ML, M LUERLOCK,, Model Number: DYNDTB5077; 17) IV ADMIN SET, 10DRP, 3 NF, 109, Model Number: DYNDTB1045; 18) DBD-IVADMISET15DRP4NEEDLEFREE VLVESTPCK1, Model Number: DYNDTB1545D; 19) IV ADMIN SET15DRPFLWCTRL1NEEDLEFREE VA, Model Number: DYNDTB2524; 20) IV ADMIN SET, 10DRP, 2NF, 96, Model Number: DYNDTB1012; 21) IV ADMIN SET, 20DRP, 2NF, 96, Model Number: DYNDTB2012; 22) IV ADMIN SET, 20DRP, NF,INJ, 96, Model Number: DYNDTB2022; 23) DBD-IV EXT SET,SMBORE EXT,8,M LUERLOCK,1, Model Number: DYNDTB5081
Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.
Feb 25, 2026
View details →Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) KIT SURG ONC INSERT CENT.VENOU, Model Number: DYKMBNDL43
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Olympus Thunderbeat 5 mm, 45 cm, Front-Actuated Grip Type S
Firm is initiating a removal due to continued reports of adverse events.
Feb 25, 2026
View details →Medline Convenience Kits: 1) DRAPE PACK-CHOICE, Model Number: DYNJ63118A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →MediHoney Gel with Active Leptospermum Honey. Model/Catalog Numbers: (1) 31805, (2) 31815. MediHoney Gel dressing with Active Leptospermum Honey provides a moist environment conducive to wound healing and is indicated for light to moderately exuding wounds.
Potential packaging failures, which could lead to a breach in the sterile barrier.
Feb 25, 2026
View details →Vicks Sinus Steam Inhaler (Vih200cv1) Recall
Recalls and safety alerts
Feb 25, 2026
View details →Estrone RIA, REF: DSL8700
The affected Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability. Only in some (very rare) cases may the issue occur.
Feb 25, 2026
View details →Medline Convenience Kits: 1) KIT ACES INSERT PEG, Model Number: DYKMBNDL130; 2) KIT ACES INSERT PEG, Model Number: DYKMBNDL130A; 3) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4; 4) NICU LINE CHANGE TRAY, Model Number: LC295; 5) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977C; 6) PICC CATHETER INSERTION TRAY, Model Number: CVI4300
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Medline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Broviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff Cat 60060
Due to damage to outer tray that can potentially compromise the sterile barrier
Feb 25, 2026
View details →Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number: DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number: DYNDH1143C; 4) BIOPSY SPECIALS PACK, Model Number: DYNDH1873; 5) BIOPSY SPECIALS PACK, Model Number: DYNDH1934; 6) BREAST BIOPSY PACK, Model Number: DYNJ17489I; 7) VOR PACK-LF, Model Number: DYNJ48539; 8) VOR PACK-LF, Model Number: DYNJ48539A; 9) IR BIOPSY PACK, Model Number: DYNJ59897B; 10) BREAST PACK, Model Number: DYNJ82998
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Feb 25, 2026
View details →Automated Impella Controller Recall
Recalls and safety alerts
Feb 25, 2026
View details →Celon ProBreath and Celon ProSleep Applicators Recall
Recalls and safety alerts
Feb 20, 2026
View details →BD Alaris™ Syringe Pumps Recall
Recalls and safety alerts
Feb 20, 2026
View details →BioFlo Dialysis Catheter With Endexo Technology Recall
Recalls and safety alerts
Feb 19, 2026
View details →Monarch Platform Recall
Recalls and safety alerts
Feb 19, 2026
View details →Teva-Indomethacin: Out of specification Recall
Recalls and safety alerts
Feb 19, 2026
View details →Bipolar Hugo RAS Systems Recall
Recalls and safety alerts
Feb 19, 2026
View details →Medline medical convenience kits packaged as: 1) POWERGLIDE DRESSING CHANGE KIT, Kit SKU DYNDH1576; 2) CENTRAL LINE TRAY W/ CHLRP SWB, Kit SKU RWM-DC57KC.
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Feb 18, 2026
View details →Kiestra ReadA Recall
Recalls and safety alerts
Feb 18, 2026
View details →Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81102RH 81104RH 81174RH 81531RH 81532RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 5mm 81107RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 2-5-2mm 81402RH 81404RH 81405RH 81474RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81473RH
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Feb 18, 2026
View details →Surepulse VS Cap Medium REF: SP-161-A1 component of VS Newborn Heart Rate Monitor
Retroactively reported; Labeling contains incorrect sizing guide for Cap, a component of the heart rate monitor.
Feb 18, 2026
View details →Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3, 11.4 Product Description: Leksell GammaPlan¿ is a computer-based system designed for Leksell Gamma Knife¿ treatment planning.
Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.
Feb 18, 2026
View details →Surepulse VS Cap Large REF: SP-162-A1 component of VS Newborn Heart Rate Monitor
Retroactively reported; Labeling contains incorrect sizing guide for Cap, a component of the heart rate monitor.
Feb 18, 2026
View details →Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;
units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.
Feb 18, 2026
View details →Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18, 2026
View details →Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav Ultrasound Catheter, Use on GE Systems 10135910RH; 10043342RH ACUSON AcuNav Ultrasound Catheter, Use on Siemens Systems 10135936RH, 08255790RH
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Feb 18, 2026
View details →CEFAZOLIN FOR INJECTION : Vials may be underfilled... Recall
Recalls and safety alerts
Feb 18, 2026
View details →Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-10-2mm 401904RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-5-2mm 401914RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-8-2-60-2-8-2mm 401932RH; Livewire Steerable Electrophysiology Catheter,8 Elec. 2-2-2mm 401652RH;
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Feb 18, 2026
View details →Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/ EZ Steer BD710DF282RRH BD710FJ282RRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710FJ282CRH;
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Feb 18, 2026
View details →Azurion 7 M12 System Model Numbers: (1)722078, (2)722223, (3)722233; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18, 2026
View details →Brilliance iCT; Product Code (REF): 728306;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Feb 18, 2026
View details →Medline 14 Fr Suction Catheter Kits SKU DYND40972 - respiratory convenience kit
A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea.
Feb 18, 2026
View details →Spectral CT; Product Code (REF): 728333;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Feb 18, 2026
View details →1) UNIVERSAL KIT, Kit SKU DT22010B; 2) VAD ACCESS KIT, Kit SKU DT22790C; 3) VAD ACCESS KIT, Kit SKU DT22790C; 4) PROCEDURE TRAY, Kit SKU DYNDA2966A; 5) PROCEDURE TRAY, Kit SKU DYNDA2966A; 6) PEDIATRIC CENTRAL LINE SMALL, Kit SKU DYNDC2208C; 7) PORT DRESSING KIT, Kit SKU DYNDC3107; 8) NON STERILE DIALYSIS KIT, Kit SKU DYNDH1029C; 9) UMBILICAL TRAY W/3.5&5FR CATH, Kit SKU UVT1250; 10) UMBILICAL VESSEL TRAY, Kit SKU UVT835; 11) HEMODIALYSIS TRAY, Kit SKU VAR101HVAA.
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Feb 18, 2026
View details →Medline medical convenience kits packaged as: STERILE RESUTURE PACK, Kit SKU DYNDL1980A
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Feb 18, 2026
View details →Medline medical convenience kits packaged as: 1) CANCER SERVICE PORT KIT, Kit SKU DT22710; 2) LINE KIT, Kit SKU DYNDA1857A; 3) PICC REMOVAL KIT, Kit SKU DYNDC1978A; 4) PORT ACCESS TRAY, Kit SKU DYNDC2367; 5) NICU CENTRAL LINE TRAY, Kit SKU DYNDC2407B; 6) CVC DRESSING CHANGE W/ TEG CHG, Kit SKU DYNDC3222A; 7) PICC PROCEDURE PACK, Kit SKU DYNJ59067A; 8) PICC PROCEDURE PACK, Kit SKU PHS950205004B.
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Feb 18, 2026
View details →Medline medical convenience kits packaged as: BLOOD CULTURE KIT NS, Kit SKU DYNDH2088
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Feb 18, 2026
View details →Medline medical convenience kits packaged as: MERCY CIRCUMCISION TRAY, Kit SKU CIT6135B
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Feb 18, 2026
View details →Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters: Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on GE Systems 10439072RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on Siemens Systems 10438577RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter, Use on Siemens Systems 10439011RH
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Feb 18, 2026
View details →Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554¿
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Feb 18, 2026
View details →Medline medical convenience kits packaged as: ADULT CATH SECUREMENT DRSG, Kit SKU DYNDC3109
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Feb 18, 2026
View details →Spectral CT 7500 on Rails; Product Code (REF): 728334;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Feb 18, 2026
View details →Lentocilin S 1200 1 Recall
Recalls and safety alerts
Feb 18, 2026
View details →Leksell Gamma Knives Recall
Recalls and safety alerts
Feb 18, 2026
View details →Raha Generator Shield Recall
Recalls and safety alerts
Feb 18, 2026
View details →Universal Viewer - System Feature Recall
Recalls and safety alerts
Feb 18, 2026
View details →Medline medical convenience kits packaged as: PEDIATRIC CATH SECUREMENT DRSG, Kit SKU DYNDC3108
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Feb 18, 2026
View details →Azurion 3 M15 System Model Numbers: (1) 722064, (2) 722222, (3) 722230, (4)722280(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18, 2026
View details →WB Impression Post Closed Tray incl. guide screw, L 13mm, TAN/POM
The mix up of the impression caps provided in the package. The provided impression caps are magenta instead of brown.
Feb 18, 2026
View details →Azurion 7 B12 System Model Numbers:(1) 722067, (2) 722225, (3) 722235; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18, 2026
View details →da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45
Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.
Feb 18, 2026
View details →Halyard TRANSPORT BAG KIT. Kit Code: LIFE0080-01.
Tyvek bag seal issue which may compromise sterility of the kit.
Feb 18, 2026
View details →ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 267-01-000-10-N UDI-DI code: 00190376541471; GMDN 12844 The ATEC Navigation Disc Prep Instruments are intended for use in surgical procedures. These instruments are non-sterile or sterile packaged, re-usable or single-use instruments designed to function with the Medtronic StealthStation System and NavLock Tracker.
Due a design issue where the navigated array connection geometry is incorrect.
Feb 18, 2026
View details →Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18, 2026
View details →IQon Spectral CT; Product Code (REF): 728332;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Feb 18, 2026
View details →Reprocessed St. Jude Supreme and Response Diagnostic Electrophysiology Catheters: St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 2-5-2mm 401211RH 401222RH 401223RH 401226RH 401227RH 401260RH 401261RH 401306RH 401309RH 401312RH 401430RH 401433RH 401435RH 401441RH 401442RH 401443RH 401448RH 401449RH 401450RH 401451RH 401474RH 401860RH 401877RH 401878RH 401890RH 401891RH 401994RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 5-5-5mm 402012RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 4 Elec. 5mm 402004RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 6 Elec. 2-5-2mm 402010RH; St. Jude Medical Supreme Diagnostic Electrophysiology Catheter 6 Elec. 5-5-5mm 402008RH 402009RH
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Feb 18, 2026
View details →Surepulse VS Cap Small REF: SP-160-A1 component of VS Newborn Heart Rate Monitor
Retroactively reported; Labeling contains incorrect sizing guide for Cap, a component of the heart rate monitor.
Feb 18, 2026
View details →Barco MNA with the HexaVue IP Integration System, Model/Catalog Number: MNA-6x0-H
This is a sub-recall of event RES 96885. The supplier identified that during their manufacturing process for a limited batch of MNA-6x0 (adapter) components the incorrect configuration file was used.
Feb 18, 2026
View details →Azurion 3 M12 System Model Numbers: (1) 722063, (2) 722221, (3) 722229; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18, 2026
View details →Medline medical convenience kits packaged as: ANESTHESIA ARTLINE PACK, Kit SKU DYNJ47387B
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Feb 18, 2026
View details →Medline medical convenience kits packaged as: ) LVAD DRESSING CHANGE TRAY, Kit SKU DM320C; 2) PORT AND IO ACCESS DRESSING KI, Kit SKU DT15780A; 3) PORT-A-CATH DRESSING CHANGE TR, Kit SKU DT16615; 4) LARGE BORE CENTRAL LINE DRSG C, Kit SKU DT19660; 5) CENTRAL LINE DRESSING CHANGE K, Kit SKU DT19665; 6) NEONATAL DRESSING CHANGE KIT, Kit SKU DT21495; 7) NICU CVC DRESSING CHANGE TRAY, Kit SKU DT22510A; 8) DRESSING CHANGE TRAY W/TEG CHG, Kit SKU DYNDC3114B; 9) DRESSING CHANGE KIT, Kit SKU DYNDC3351.
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Feb 18, 2026
View details →Surepulse VS Cap Extra Small REF: SP-159-A1 component of VS Newborn Heart Rate Monitor
Retroactively reported; Labeling contains incorrect sizing guide for Cap, a component of the heart rate monitor.
Feb 18, 2026
View details →ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 266-01-000-10-N UDI-DI code: 00190376476179; GMDN 12844 The ATEC Navigation Disc Prep Instruments are intended for use in surgical procedures. These instruments are non-sterile or sterile packaged, re-usable or single-use instruments designed to function with the Medtronic StealthStation System and NavLock Tracker.
Due a design issue where the navigated array connection geometry is incorrect.
Feb 18, 2026
View details →Medline medical convenience kits packaged as: SUTURE REMOVAL TRAY, Kit SKU DYNDR1277A
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Feb 18, 2026
View details →Azurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18, 2026
View details →Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11.1 Product Description: Leksell GammaPlan¿ is a computer-based system designed for Leksell Gamma Knife¿ treatment planning.
Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.
Feb 18, 2026
View details →Intubation Tray (Adult) containing Suction Catheter Kits SKU DYNJAA245B
A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea.
Feb 18, 2026
View details →Azurion 5 M20 System Model Numbers: (1)722228, (2)722232, (3)722281(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Feb 18, 2026
View details →Surepulse VS Cap Extra Large REF: SP-163-A1 component of VS Newborn Heart Rate Monitor
Retroactively reported; Labeling contains incorrect sizing guide for Cap, a component of the heart rate monitor.
Feb 18, 2026
View details →Brevera Breast Biopsy System Disposable 9 Gauge Needle. Biopsy Instrument.
Presence of particulates in affected devices that can be deposited into breast tissue during use.
Feb 18, 2026
View details →Erbe Flexible Cryoprobe Recall
Recalls and safety alerts
Feb 17, 2026
View details →APO-CANDESARTAN: Affected lots were manufactured u... Recall
Recalls and safety alerts
Feb 16, 2026
View details →Trilogy EVO and EV300 Recall
Recalls and safety alerts
Feb 13, 2026
View details →Medline Kits containing Surgical Gowns Recall
Recalls and safety alerts
Feb 11, 2026
View details →Medihoney® Wound and Burn Products Recall
Recalls and safety alerts
Feb 11, 2026
View details →VISERA™ Imaging System Recall
Recalls and safety alerts
Feb 10, 2026
View details →Alphenix Systems Recall
Recalls and safety alerts
Feb 10, 2026
View details →TARO-CANDESARTAN: Affected lots were manufactured ... Recall
Recalls and safety alerts
Feb 9, 2026
View details →Cefazolin for Injection: Affected lots are package... Recall
Recalls and safety alerts
Feb 9, 2026
View details →MAR-AMLODIPINE: affected lots may contain tablet(s... Recall
Recalls and safety alerts
Feb 9, 2026
View details →Two lots of MAR-Amlodipine Recall
Recalls and safety alerts
Feb 7, 2026
View details →UHI-3 High Flow Insufflation System - Insufflator Recall
Recalls and safety alerts
Feb 6, 2026
View details →Philips IntelliVue Patient Monitors Recall
Recalls and safety alerts
Nov 18, 2025