Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYST , Catalog Number: PB41RHRM3; 2) KIT, ROBOTIC HYST , Catalog Number: PB41RHRM4; 3) KIT, C SECTION , Catalog Number: PLVMCS22G; 4) KIT, C SECTION MARTIN , Catalog Number: PL22CSMMI; 5) KIT, C SECTION MARTIN , Catalog Number: PL22CSMMI; 6) KIT, C SECTION OREGON HEALTH , Catalog Number: PL57CS60N; 7) C SECTION PACK , Catalog Number: SMAHBCSECB; 8) C SECTION PACK , Catalog Number: SMAHBCSEC3
Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYST , Catalog Number: PB41RHRM3; 2) KIT, ROBOTIC HYST , Catalog Number: PB41RHRM4; 3) KIT, C SECTION , Catalog Number: PLVMCS22G; 4) KIT, C SECTION MARTIN , Catalog Number: PL22CSMMI; 5) KIT, C SECTION MARTIN , Catalog Number: PL22CSMMI; 6) KIT, C SECTION OREGON HEALTH , Catalog Number: PL57CS60N; 7) C SECTION PACK , Catalog Number: SMAHBCSECB; 8) C SECTION PACK , Catalog Number: SMAHBCSEC3 was recalled because the urinary catheter trays contain cliniclean pericare wipes which may potentially be contaminated with a lubricant used during the manufacturing process. if a contaminated wipe is used on a patient, localized skin effects may occur. these may include irritation, erythema, or mild inflammation.. If you have this product, stop using it until you verify your lot or model details.
By RecallRadar Editorial Team
Sourced from FDA. Last verified September 15, 2026.
On this page
Free checker
Check if you qualify
See whether your device may qualify for a free repair or replacement program before you spend money.
Check if you qualifyKey facts
- Date announced
- September 9, 2026
- Source agency
- FDA
This recall involves safety concerns. Take action promptly.
✨ Plain-English Summary
Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, ROBOTIC HYST , Catalog Number: PB41RHRM3; 2) KIT, ROBOTIC HYST , Catalog Number: PB41RHRM4; 3) KIT, C SECTION , Catalog Number: PLVMCS22G; 4) KIT, C SECTION MARTIN , Catalog Number: PL22CSMMI; 5) KIT, C SECTION MARTIN , Catalog Number: PL22CSMMI; 6) KIT, C SECTION OREGON HEALTH , Catalog Number: PL57CS60N; 7) C SECTION PACK , Catalog Number: SMAHBCSECB; 8) C SECTION PACK , Catalog Number: SMAHBCSEC3 was recalled because the urinary catheter trays contain cliniclean pericare wipes which may potentially be contaminated with a lubricant used during the manufacturing process. if a contaminated wipe is used on a patient, localized skin effects may occur. these may include irritation, erythema, or mild inflammation.. If you have this product, stop using it until you verify your lot or model details.
Is This Recall Still Active?
Status: Yes - Active
This recall is currently active. If your device matches the affected models, follow the official instructions as soon as possible.
What's Being Recalled
Quoted source excerpt from the official notice.
The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation.
Quick Info
Key facts from the official notice, kept in the article flow for easy cross-checking.
- Brand
- Cardinal health 200, llc
- Category
- medical
- Source agency
- FDA
- Announced
- September 9, 2026
- Country
- US
- Status
- Active
Who's Affected
Specific models and identifiers listed in the source notice:
Model #1
1) PB41RHRM3, UDI-DI: 10198956294219(each), 50198956294217(case), Lot Number: NGKU3049 ; 2) PB41RHRM3, UDI-DI: 10198956294219(each), 50198956294217(case), Lot Number: NGKU4043 ; 3) PB41RHRM4, UDI-DI: 10197106968413(each), 50197106968411(case), Lot Number: NGKU4716 ; 4) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0957 ; 5) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0856 ; 6) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0957 ; 7) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU2052 ; 8) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU2052 ; 9) PL22CSMMI, UDI-DI: 10195594952600(each), 50195594952608(case), Lot Number: NGKU0129 ; 10) PL22CSM13, UDI-DI: 10198956457553(each), 50198956457551(case), Lot Number: NGKU0129 ; 11) PL57CS60N, UDI-DI: 10197106889848(each), 50197106889846(case), Lot Number: NGKU2052 ; 12) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: NGKU4716 ; 13) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: NGKU4716 ; 14) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: 20251217 ; 15) SMAHBCSEC3, UDI-DI: 10198956267183(each), 50198956267181(case), Lot Number: NGKU2887
| # | Model |
|---|---|
| 1 | 1) PB41RHRM3, UDI-DI: 10198956294219(each), 50198956294217(case), Lot Number: NGKU3049 ; 2) PB41RHRM3, UDI-DI: 10198956294219(each), 50198956294217(case), Lot Number: NGKU4043 ; 3) PB41RHRM4, UDI-DI: 10197106968413(each), 50197106968411(case), Lot Number: NGKU4716 ; 4) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0957 ; 5) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0856 ; 6) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0957 ; 7) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU2052 ; 8) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU2052 ; 9) PL22CSMMI, UDI-DI: 10195594952600(each), 50195594952608(case), Lot Number: NGKU0129 ; 10) PL22CSM13, UDI-DI: 10198956457553(each), 50198956457551(case), Lot Number: NGKU0129 ; 11) PL57CS60N, UDI-DI: 10197106889848(each), 50197106889846(case), Lot Number: NGKU2052 ; 12) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: NGKU4716 ; 13) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: NGKU4716 ; 14) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: 20251217 ; 15) SMAHBCSEC3, UDI-DI: 10198956267183(each), 50198956267181(case), Lot Number: NGKU2887 |
Technical model codes are manufacturer identifiers. Match both the product name and model code from your device settings, label, or packaging.
What You Should Do
Follow the remedy path from the official notice using the exact model and serial details above.
- 1Stop using the affected product until you verify whether your item is impacted.
- 2Check the official recall notice and match your model, lot, or serial details.
- 3Follow the official instructions for repair, replacement, refund, or disposal.
Source & Verification
This page summarizes the official notice and keeps key facts easy to verify.
- Official source
- FDA
- Announced by source
- September 9, 2026
- Last reviewed by RecallRadar
- September 15, 2026
Frequently Asked Questions
Am I affected by this recall?
Check the affected models section on this page. The official notice currently lists 1) PB41RHRM3, UDI-DI: 10198956294219(each), 50198956294217(case), Lot Number: NGKU3049 ; 2) PB41RHRM3, UDI-DI: 10198956294219(each), 50198956294217(case), Lot Number: NGKU4043 ; 3) PB41RHRM4, UDI-DI: 10197106968413(each), 50197106968411(case), Lot Number: NGKU4716 ; 4) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0957 ; 5) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0856 ; 6) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU0957 ; 7) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU2052 ; 8) PLVMCS22G, UDI-DI: 10198956198890(each), 50198956198898(case), Lot Number: NGKU2052 ; 9) PL22CSMMI, UDI-DI: 10195594952600(each), 50195594952608(case), Lot Number: NGKU0129 ; 10) PL22CSM13, UDI-DI: 10198956457553(each), 50198956457551(case), Lot Number: NGKU0129 ; 11) PL57CS60N, UDI-DI: 10197106889848(each), 50197106889846(case), Lot Number: NGKU2052 ; 12) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: NGKU4716 ; 13) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: NGKU4716 ; 14) SMAHBCSECB, UDI-DI: 10197106958926(each), 50197106958924(case), Lot Number: 20251217 ; 15) SMAHBCSEC3, UDI-DI: 10198956267183(each), 50198956267181(case), Lot Number: NGKU2887.
What is the remedy for this recall?
Follow the official instructions for repair, replacement, refund, or disposal on the cardinal health 200, llc recall page.
Where can I verify recall details?
Use the official source link in this page to confirm dates, eligibility, and next steps: https://api.fda.gov/device/enforcement.json?search=recall_number:Z-3102-2026&limit=1
Keep browsing with these internal hubs: cardinal health 200, llc recall hub and Medical recall hub.
Official Resources
Use these primary sources for final eligibility and claim instructions.
View official notice from FDASee an inaccuracy?
Help us improve the accuracy of this recall information.
On this page
Free checker
Check if you qualify
See whether your device may qualify for a free repair or replacement program before you spend money.
Check if you qualifyKey facts
- Date announced
- September 9, 2026
- Source agency
- FDA
Related Recalls
GRANUFLO Dry Acid Concentrate: Product Numbers: 0FD2251-3B; 0FD3251-3B; Dry Acid Concentrate for Bicarbonate Hemodialysis
Complaints of blue-green discoloration in product
Sep 9, 2026
5008X HD/HDF with Twister Blood Tubing Set Part Number 03-5350-8
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Sep 9, 2026
Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) PBCGLVBLI, Catalog Number: KIT, BLO LAVH ; 2) PB15HYFCY, Catalog Number: KIT, UMMC AB HYST II 170032 ; 3) SMA4FTLSCA, Catalog Number: TLH PACK
The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation.
Sep 9, 2026
Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT, GYN ROBOTIC GSAM , Catalog Number: PBCCGRSGD; 2) KIT, GYN ROBOT ACCESSORY , Catalog Number: PB35GYBJC; 3) KIT, PRIMARY RESECTION , Catalog Number: PGVMOTBB19; 4) KIT, LIVER TRANSPLANT SUPPORT , Catalog Number: PGVM07C26; 5) KIT, LIVER TRANSPLANT SUPPORT , Catalog Number: PGVM07C26; 6) KIT, LIVER TRANSPLANT SUPPORT , Catalog Number: PGV2LTNHN; 7) KIT,UMMC GENERAL MAJOR TRANSPL, Catalog Number: PG15OT92A; 8) KIT,MAJOR LITHOTOMY NEW HAN , Catalog Number: PG23MLNLD; 9) KIT, GENDER CARE , Catalog Number: PG29GCPCK; 10) KIT, GENDER CARE , Catalog Number: PG29GCPC3; 11) KIT, MUH LIVER , Catalog Number: PG33LVMNC; 12) KIT, MIS HERNIA , Catalog Number: PG35HPBJK; 13) KIT, GENERAL MAJOR , Catalog Number: PG43MJEJD; 14) KIT, OPEN THORACOTOMY CMC , Catalog Number: PVCGOTACF; 15) KIT, MER THORACOTOMY ABDOMINAL, Catalog Number: PVX4TAMYB; 16) KIT, MER THORACOTOMY ABDOMINAL, Catalog Number: PVX4TAMYC; 17) KIT, HEART II , Catalog Number: PV23HEVAE; 18) KIT, FLAP ACCESSORY , Catalog Number: PW35FABJE; 19) HMC ROBOTIC PACK , Catalog Number: SBA23RBHNG; 20) HMC ROBOTIC PACK , Catalog Number: SBA23RBHNH
The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation.
Sep 9, 2026
Velys Enabling Tech Robotic-Assist Station REF: 451611001
Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.
Sep 9, 2026