Olympus Resection Sheath, 28 Fr.; Model No. A22043A. Resection sheath for urologic applications.
Olympus Resection Sheath, 28 Fr.; Model No. A22043A. Resection sheath for urologic applications. was recalled because complaints of the ceramic tip of the resection sheath breaking have been received.. If you have this product, stop using it until you verify your lot or model details.
By RecallRadar Editorial Team
Sourced from FDA. Last verified March 15, 2026.
On this page
Free checker
Check if you qualify
See whether your device may qualify for a free repair or replacement program before you spend money.
Check if you qualifyKey facts
- Date announced
- March 4, 2026
- Source agency
- FDA
This recall involves safety concerns. Take action promptly.
✨ Plain-English Summary
Olympus Resection Sheath, 28 Fr.; Model No. A22043A. Resection sheath for urologic applications. was recalled because complaints of the ceramic tip of the resection sheath breaking have been received.. If you have this product, stop using it until you verify your lot or model details.
Is This Recall Still Active?
Status: Yes - Active
This recall is currently active. If your device matches the affected models, follow the official instructions as soon as possible.
What's Being Recalled
Quoted source excerpt from the official notice.
Complaints of the ceramic tip of the resection sheath breaking have been received.
Quick Info
Key facts from the official notice, kept in the article flow for easy cross-checking.
- Brand
- Olympus corporation of the americas
- Category
- medical
- Source agency
- FDA
- Announced
- March 4, 2026
- Country
- US
- Status
- Active
Who's Affected
Specific models and identifiers listed in the source notice:
Model #1
Model No. A22043A; UDI: 04042761021005; All Lots.
| # | Model |
|---|---|
| 1 | Model No. A22043A; UDI: 04042761021005; All Lots. |
Technical model codes are manufacturer identifiers. Match both the product name and model code from your device settings, label, or packaging.
What You Should Do
Follow the remedy path from the official notice using the exact model and serial details above.
- 1Stop using the affected product until you verify whether your item is impacted.
- 2Check the official recall notice and match your model, lot, or serial details.
- 3Follow the official instructions for repair, replacement, refund, or disposal.
Source & Verification
This page summarizes the official notice and keeps key facts easy to verify.
- Official source
- FDA
- Announced by source
- March 4, 2026
- Last reviewed by RecallRadar
- March 15, 2026
Frequently Asked Questions
Am I affected by this recall?
Check the affected models section on this page. The official notice currently lists Model No. A22043A; UDI: 04042761021005; All Lots..
What is the remedy for this recall?
Follow the official instructions for repair, replacement, refund, or disposal on the olympus corporation of the americas recall page.
Where can I verify recall details?
Use the official source link in this page to confirm dates, eligibility, and next steps: https://api.fda.gov/device/enforcement.json?search=recall_number:Z-1445-2026&limit=1
Keep browsing with these internal hubs: olympus corporation of the americas recall hub and Medical recall hub.
Official Resources
Use these primary sources for final eligibility and claim instructions.
View official notice from FDASee an inaccuracy?
Help us improve the accuracy of this recall information.
On this page
Free checker
Check if you qualify
See whether your device may qualify for a free repair or replacement program before you spend money.
Check if you qualifyKey facts
- Date announced
- March 4, 2026
- Source agency
- FDA
Related Recalls
Rapid Refill Continuous Injection System, UPN M00566001
Devices contain recalled Medline Namic Angiographic Control Syringe with Rotating Adaptor. There is a risk of syringe loosening or disconnecting from the manifold, resulting in loose connection or complete disconnection. May lead to difficulty connecting syringe, inability to aspirate or inject properly, fluid leakage, or separation of the syringe from the system, potentially requiring device exchange and prolonging the procedure.
Jul 22, 2026
Papablic Replacement Brush Heads. Model Number: Papablic18B
Potential for brush heads to break.
Jul 22, 2026
Orthofix MINIRAIL, STANDARD LENGTHENER Model/Catalog Number: M101, Unilateral orthopedic/trauma external fixation device
Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122 was labeled as M111
Jul 22, 2026
Papablic Replacement Brush Heads. Model Number: PAPA317AA.
Potential for brush heads to break.
Jul 22, 2026
ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Jul 22, 2026