Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 EBSI1596 EBSI1781 EBSI1812 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845
Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 EBSI1596 EBSI1781 EBSI1812 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845 was recalled because affected lots may exhibit issues with the silicone seal on tego connectors included in kits. issues include silicone seal doming and silicone seal tearing. these issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.. If you have this product, stop using it until you verify your lot or model details.
By RecallRadar Editorial Team
Sourced from FDA. Last verified March 15, 2026.
On this page
Free checker
Check if you qualify
See whether your device may qualify for a free repair or replacement program before you spend money.
Check if you qualifyKey facts
- Date announced
- March 4, 2026
- Source agency
- FDA
This recall involves safety concerns. Take action promptly.
✨ Plain-English Summary
Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 EBSI1596 EBSI1781 EBSI1812 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845 was recalled because affected lots may exhibit issues with the silicone seal on tego connectors included in kits. issues include silicone seal doming and silicone seal tearing. these issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.. If you have this product, stop using it until you verify your lot or model details.
Is This Recall Still Active?
Status: Yes - Active
This recall is currently active. If your device matches the affected models, follow the official instructions as soon as possible.
What's Being Recalled
Quoted source excerpt from the official notice.
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Quick Info
Key facts from the official notice, kept in the article flow for easy cross-checking.
- Brand
- Medline industries, lp
- Category
- medical
- Source agency
- FDA
- Announced
- March 4, 2026
- Country
- US
- Status
- Active
Who's Affected
Specific models and identifiers listed in the source notice:
Model #1
Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 UDI-DI (ea) 10193489111538 UDI-DI (case) 20193489111535 Kit lots 2024052090 2024072690 2024100390 2024121290 2025020490 2025041590 2025062790 2025090890 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 UDI-DI (ea) 10889942723508 UDI-DI (case) 40889942723509 Kit lots 2024051690 2024073190 2024092490 2024121190 2025020790 2025032790 2025052990 EBSI1596 UDI-DI (ea) 10653160357162 UDI-DI (case) 00653160357165 Kit lots 2024070190 2024082390 2024102390 2024121790 2025031090 2025040190 2025052790 2025072890 EBSI1781 UDI-DI (ea) 10653160379638 UDI-DI (case) 00653160379631 Kit lots 2024100490 2024123090 EBSI1812 UDI-DI (ea) 10653160386735 UDI-DI (case) 00653160386738 Kit lots 2025032190 2025052990 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 UDI-DI (ea) 10653160345343 UDI-DI (case) 00653160345346 Kit lots 2024050680 2024072980 2024090380 2024111290 2024121190 2025032890 2025061790 2025072390 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 UDI-DI (ea) 10193489199307 UDI-DI (case) 20193489199304 Kit lots 2024052080 2024070880 2024110490 2025010290 2025022090 2025030390 2025040990 2025072990 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845 UDI-DI (ea) 10653160355151 UDI-DI (case) 00653160355154 Kit lots 2024060790 2024080790 2024100990 2024112590 2025042490 2025082090
| # | Model |
|---|---|
| 1 | Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 UDI-DI (ea) 10193489111538 UDI-DI (case) 20193489111535 Kit lots 2024052090 2024072690 2024100390 2024121290 2025020490 2025041590 2025062790 2025090890 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 UDI-DI (ea) 10889942723508 UDI-DI (case) 40889942723509 Kit lots 2024051690 2024073190 2024092490 2024121190 2025020790 2025032790 2025052990 EBSI1596 UDI-DI (ea) 10653160357162 UDI-DI (case) 00653160357165 Kit lots 2024070190 2024082390 2024102390 2024121790 2025031090 2025040190 2025052790 2025072890 EBSI1781 UDI-DI (ea) 10653160379638 UDI-DI (case) 00653160379631 Kit lots 2024100490 2024123090 EBSI1812 UDI-DI (ea) 10653160386735 UDI-DI (case) 00653160386738 Kit lots 2025032190 2025052990 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 UDI-DI (ea) 10653160345343 UDI-DI (case) 00653160345346 Kit lots 2024050680 2024072980 2024090380 2024111290 2024121190 2025032890 2025061790 2025072390 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 UDI-DI (ea) 10193489199307 UDI-DI (case) 20193489199304 Kit lots 2024052080 2024070880 2024110490 2025010290 2025022090 2025030390 2025040990 2025072990 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845 UDI-DI (ea) 10653160355151 UDI-DI (case) 00653160355154 Kit lots 2024060790 2024080790 2024100990 2024112590 2025042490 2025082090 |
Technical model codes are manufacturer identifiers. Match both the product name and model code from your device settings, label, or packaging.
What You Should Do
Follow the remedy path from the official notice using the exact model and serial details above.
- 1Stop using the affected product until you verify whether your item is impacted.
- 2Check the official recall notice and match your model, lot, or serial details.
- 3Follow the official instructions for repair, replacement, refund, or disposal.
Source & Verification
This page summarizes the official notice and keeps key facts easy to verify.
- Official source
- FDA
- Announced by source
- March 4, 2026
- Last reviewed by RecallRadar
- March 15, 2026
Frequently Asked Questions
Am I affected by this recall?
Check the affected models section on this page. The official notice currently lists Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 UDI-DI (ea) 10193489111538 UDI-DI (case) 20193489111535 Kit lots 2024052090 2024072690 2024100390 2024121290 2025020490 2025041590 2025062790 2025090890 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 UDI-DI (ea) 10889942723508 UDI-DI (case) 40889942723509 Kit lots 2024051690 2024073190 2024092490 2024121190 2025020790 2025032790 2025052990 EBSI1596 UDI-DI (ea) 10653160357162 UDI-DI (case) 00653160357165 Kit lots 2024070190 2024082390 2024102390 2024121790 2025031090 2025040190 2025052790 2025072890 EBSI1781 UDI-DI (ea) 10653160379638 UDI-DI (case) 00653160379631 Kit lots 2024100490 2024123090 EBSI1812 UDI-DI (ea) 10653160386735 UDI-DI (case) 00653160386738 Kit lots 2025032190 2025052990 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 UDI-DI (ea) 10653160345343 UDI-DI (case) 00653160345346 Kit lots 2024050680 2024072980 2024090380 2024111290 2024121190 2025032890 2025061790 2025072390 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 UDI-DI (ea) 10193489199307 UDI-DI (case) 20193489199304 Kit lots 2024052080 2024070880 2024110490 2025010290 2025022090 2025030390 2025040990 2025072990 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845 UDI-DI (ea) 10653160355151 UDI-DI (case) 00653160355154 Kit lots 2024060790 2024080790 2024100990 2024112590 2025042490 2025082090.
What is the remedy for this recall?
Follow the official instructions for repair, replacement, refund, or disposal on the medline industries, lp recall page.
Where can I verify recall details?
Use the official source link in this page to confirm dates, eligibility, and next steps: https://api.fda.gov/device/enforcement.json?search=recall_number:Z-1467-2026&limit=1
Keep browsing with these internal hubs: medline industries, lp recall hub and Medical recall hub.
Official Resources
Use these primary sources for final eligibility and claim instructions.
View official notice from FDASee an inaccuracy?
Help us improve the accuracy of this recall information.
On this page
Free checker
Check if you qualify
See whether your device may qualify for a free repair or replacement program before you spend money.
Check if you qualifyKey facts
- Date announced
- March 4, 2026
- Source agency
- FDA
Related Recalls
Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electrical stimulator, a TCS stimulator, a visual stimulator and an auditory stimulator. The low current stimulator has one (1) constant current output pair with a current range of 0.01-20 mA. The TCS stimulator has nine (9) independent constant current/constant voltage outputs, with current range of 50-1500 mA and voltage range of 50-1000 V. Component: Is the cortical component of the IOMAX system
Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.
Oct 21, 2026
JenaValve Trilogy Delivery Catheter, Size 23MM REF: TRILOGY-DC-S
Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.
Oct 21, 2026
Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone matrix (DBM) allograft product having a particular physical form and/or handling property. Grafton DBM Gel and Grafton Plus DBM Paste are produced from a powder form of DBM, while Grafton DBM Flex, Putty, Matrix, Crunch" and Orthoblend are produced from a fiber form of DBM Crunch" also contains demineralized bone chips/cubes, while Grafton DBM Orthoblend contains non-demineralized cancellous bone chips in addition to DBM. Grafton DBM and Grafton Plus DBM are malleable and can be molded or cut into various sizes and shapes according to the intended implant site.
Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection.
Oct 21, 2026
Medline surgical convenience kits labeled as: 1) OPEN HEART CDS, Medline product number/SKU CDS840094R; 2) OPEN HEART CDS, Medline product number/SKU CDS840171AG; 3) MHC AV FISTULA, Medline product number/SKU DYNJ903705K; 4) PERIPHERAL VASC MAJR-LF, Medline product number/SKU CDS840379O; 5) MINOR VASCULAR PACK-LF, Medline product number/SKU DYNJ0349727D; 6) SJHC VASCULAR PK-LF, Medline product number/SKU DYNJ19906L; 7) CABG PACK-LF, Medline product number/SKU DYNJ0824648K; 8) PERIPHERAL VASC MNR-LF, Medline product number/SKU DYNJ22721N; 9) HEART A PACK, Medline product number/SKU DYNJ0362405S; 10) OPEN HEART, Medline product number/SKU DYNJ907676I; 11) FEM POP MARYVIEW, Medline product number/SKU DYNJ911714; 12) CABG PACK, Medline product number/SKU DYNJ906346G; 13) DEBOER HEART MAIN, Medline product number/SKU DYNJ907914I; 14) OPEN HEART A & B PACK, Medline product number/SKU DYNJ9718008AU; 15) OPEN HEART, Medline product number/SKU DYNJ911483A; 16) CABG ACCESSORY PACK, Medline product number/SKU DYNJ59265; 17) HEART, Medline product number/SKU DYNJ908779B; 18) CARDIOVASCULAR PACK, Medline product number/SKU DYNJ69507F; 19) OFF PUMP CABG CDS, Medline product number/SKU CDS840087AN; 20) PACK CABG CHRG, Medline product number/SKU DYNJ905683D; 21) HEART CDS, Medline product number/SKU CDS840179N;
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Oct 21, 2026
Medline convenience kit labeled as: FOOT ANKLE. Medline Kit SKU DYNJ65395A
Discrepancy in the shelf-life of a component
Oct 21, 2026