Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-026, 2088026-115, 2088026-043, K2042VJED, 2088026-132, 2088026-003, 2088026-110, 2089629-003, 2066908-136, 2066908-150
Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-026, 2088026-115, 2088026-043, K2042VJED, 2088026-132, 2088026-003, 2088026-110, 2089629-003, 2066908-136, 2066908-150 was recalled because there is a potential cybersecurity vulnerability affecting certain versions of centricity universal viewer. user login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.. If you have this product, stop using it until you verify your lot or model details.
By RecallRadar Editorial Team
Sourced from FDA. Last verified April 8, 2026.
On this page
Free checker
Check if you qualify
See whether your device may qualify for a free repair or replacement program before you spend money.
Check if you qualifyKey facts
- Date announced
- March 25, 2026
- Source agency
- FDA
This recall involves safety concerns. Take action promptly.
✨ Plain-English Summary
Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-026, 2088026-115, 2088026-043, K2042VJED, 2088026-132, 2088026-003, 2088026-110, 2089629-003, 2066908-136, 2066908-150 was recalled because there is a potential cybersecurity vulnerability affecting certain versions of centricity universal viewer. user login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.. If you have this product, stop using it until you verify your lot or model details.
Is This Recall Still Active?
Status: Yes - Active
This recall is currently active. If your device matches the affected models, follow the official instructions as soon as possible.
What's Being Recalled
Quoted source excerpt from the official notice.
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
Quick Info
Key facts from the official notice, kept in the article flow for easy cross-checking.
- Brand
- Ge medical systems, llc
- Category
- medical
- Source agency
- FDA
- Announced
- March 25, 2026
- Country
- US
- Status
- Active
Who's Affected
Specific models and identifiers listed in the source notice:
Model #1
Product ID Numbers: 4446-2-Centricity Universal Viewer-00558643 100256-3-Centricity Universal Viewer-00558700 101044-1-Centricity Universal Viewer-00558694 13240-2-Centricity Universal Viewer-00558103 4441-1-Centricity Universal Viewer-00558677 4578-1-Centricity Universal Viewer-00558340 10831-2-Centricity Universal Viewer-00558682 656341-Centricity Universal Viewer-01431499 50567-Centricity Universal Viewer-02176369 2690973-Centricity Universal Viewer-01454622 EG1526-Centricity Universal Viewer-00158976 A91901-Centricity Universal Viewer-00044549 M3488569-Centricity Universal Viewer-00071707 25540620 - HC4105-Centricity Universal Viewer-00001393 416005-Centricity Universal Viewer-00260808 1868708X - 556162-Centricity Universal Viewer-00157857 HKOB01-Centricity Universal Viewer-01282268 BMH-Centricity Universal Viewer-01820534 JIPMER-Centricity Universal Viewer-01820555 4347967-Centricity Universal Viewer-01474540 651342-Centricity Universal Viewer-01365218 651342-Centricity Universal Viewer-01365213 A3431341-Centricity Universal Viewer-00545677 473265-Centricity Universal Viewer-01442014 JP-131575-Centricity Universal Viewer-01275307 442900-Centricity Universal Viewer-01460668 577127-Centricity Universal Viewer-01444349 349551-Centricity Universal Viewer-02350518 PL3277-Centricity Universal Viewer-00038831 RU3866-Centricity Universal Viewer-01445731 RU8945-Centricity Universal Viewer-01445740 RU2580-Centricity Universal Viewer-01445743 1354732-Centricity Universal Viewer-01273292 RU3453-Centricity Universal Viewer-01273249 SA2160-Centricity Universal Viewer-00111827 SA1208-Centricity Universal Viewer-00066587 882983-Centricity Universal Viewer-02277629 ZA2533-Centricity Universal Viewer-00116577
| # | Model |
|---|---|
| 1 | Product ID Numbers: 4446-2-Centricity Universal Viewer-00558643 100256-3-Centricity Universal Viewer-00558700 101044-1-Centricity Universal Viewer-00558694 13240-2-Centricity Universal Viewer-00558103 4441-1-Centricity Universal Viewer-00558677 4578-1-Centricity Universal Viewer-00558340 10831-2-Centricity Universal Viewer-00558682 656341-Centricity Universal Viewer-01431499 50567-Centricity Universal Viewer-02176369 2690973-Centricity Universal Viewer-01454622 EG1526-Centricity Universal Viewer-00158976 A91901-Centricity Universal Viewer-00044549 M3488569-Centricity Universal Viewer-00071707 25540620 - HC4105-Centricity Universal Viewer-00001393 416005-Centricity Universal Viewer-00260808 1868708X - 556162-Centricity Universal Viewer-00157857 HKOB01-Centricity Universal Viewer-01282268 BMH-Centricity Universal Viewer-01820534 JIPMER-Centricity Universal Viewer-01820555 4347967-Centricity Universal Viewer-01474540 651342-Centricity Universal Viewer-01365218 651342-Centricity Universal Viewer-01365213 A3431341-Centricity Universal Viewer-00545677 473265-Centricity Universal Viewer-01442014 JP-131575-Centricity Universal Viewer-01275307 442900-Centricity Universal Viewer-01460668 577127-Centricity Universal Viewer-01444349 349551-Centricity Universal Viewer-02350518 PL3277-Centricity Universal Viewer-00038831 RU3866-Centricity Universal Viewer-01445731 RU8945-Centricity Universal Viewer-01445740 RU2580-Centricity Universal Viewer-01445743 1354732-Centricity Universal Viewer-01273292 RU3453-Centricity Universal Viewer-01273249 SA2160-Centricity Universal Viewer-00111827 SA1208-Centricity Universal Viewer-00066587 882983-Centricity Universal Viewer-02277629 ZA2533-Centricity Universal Viewer-00116577 |
Technical model codes are manufacturer identifiers. Match both the product name and model code from your device settings, label, or packaging.
What You Should Do
Follow the remedy path from the official notice using the exact model and serial details above.
- 1Stop using the affected product until you verify whether your item is impacted.
- 2Check the official recall notice and match your model, lot, or serial details.
- 3Follow the official instructions for repair, replacement, refund, or disposal.
Source & Verification
This page summarizes the official notice and keeps key facts easy to verify.
- Official source
- FDA
- Announced by source
- March 25, 2026
- Last reviewed by RecallRadar
- April 8, 2026
Frequently Asked Questions
Am I affected by this recall?
Check the affected models section on this page. The official notice currently lists Product ID Numbers: 4446-2-Centricity Universal Viewer-00558643 100256-3-Centricity Universal Viewer-00558700 101044-1-Centricity Universal Viewer-00558694 13240-2-Centricity Universal Viewer-00558103 4441-1-Centricity Universal Viewer-00558677 4578-1-Centricity Universal Viewer-00558340 10831-2-Centricity Universal Viewer-00558682 656341-Centricity Universal Viewer-01431499 50567-Centricity Universal Viewer-02176369 2690973-Centricity Universal Viewer-01454622 EG1526-Centricity Universal Viewer-00158976 A91901-Centricity Universal Viewer-00044549 M3488569-Centricity Universal Viewer-00071707 25540620 - HC4105-Centricity Universal Viewer-00001393 416005-Centricity Universal Viewer-00260808 1868708X - 556162-Centricity Universal Viewer-00157857 HKOB01-Centricity Universal Viewer-01282268 BMH-Centricity Universal Viewer-01820534 JIPMER-Centricity Universal Viewer-01820555 4347967-Centricity Universal Viewer-01474540 651342-Centricity Universal Viewer-01365218 651342-Centricity Universal Viewer-01365213 A3431341-Centricity Universal Viewer-00545677 473265-Centricity Universal Viewer-01442014 JP-131575-Centricity Universal Viewer-01275307 442900-Centricity Universal Viewer-01460668 577127-Centricity Universal Viewer-01444349 349551-Centricity Universal Viewer-02350518 PL3277-Centricity Universal Viewer-00038831 RU3866-Centricity Universal Viewer-01445731 RU8945-Centricity Universal Viewer-01445740 RU2580-Centricity Universal Viewer-01445743 1354732-Centricity Universal Viewer-01273292 RU3453-Centricity Universal Viewer-01273249 SA2160-Centricity Universal Viewer-00111827 SA1208-Centricity Universal Viewer-00066587 882983-Centricity Universal Viewer-02277629 ZA2533-Centricity Universal Viewer-00116577.
What is the remedy for this recall?
Follow the official instructions for repair, replacement, refund, or disposal on the ge medical systems, llc recall page.
Where can I verify recall details?
Use the official source link in this page to confirm dates, eligibility, and next steps: https://api.fda.gov/device/enforcement.json?search=recall_number:Z-1571-2026&limit=1
Keep browsing with these internal hubs: ge medical systems, llc recall hub and Medical recall hub.
Official Resources
Use these primary sources for final eligibility and claim instructions.
View official notice from FDASee an inaccuracy?
Help us improve the accuracy of this recall information.
On this page
Free checker
Check if you qualify
See whether your device may qualify for a free repair or replacement program before you spend money.
Check if you qualifyKey facts
- Date announced
- March 25, 2026
- Source agency
- FDA
Related Recalls
XTANT MEDICAL Xpress Pedicle Screw System Dual Lead 6.5 x 50 mm, Part Number X073-6555-DL.
Units from one lot were mislabeled with an incorrect part number, identifying the device as a shorter screw length than it physically is.
Apr 15, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
See Luer Cap Set, MPC-130, set, administration, intravascular
Molded Products, Inc. is voluntarily conducting a medical device recall of the MPC-130 See Luer Cap Set based on a confirmed complaint of the threaded sleeve not being engaged and becoming un-attached.
Apr 15, 2026
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. NON VASCULAR PACK, Medline SKU # DYNJ68037C; 2. VEIN, Medline SKU # DYNJ89845; 3. VEIN, Medline SKU # DYNJ89845A.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Apr 15, 2026
COOK MEDICAL Blue Rhino¿ G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703; Reference Part Number C-PTISY-100-UNL-HC-G-NA, Order Number G57706; Reference Part Number C-PTISY-100-UNS-HC-G-NA, Order Number G57704.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15, 2026