brahms gmbhmedicalHigh PriorityActive

B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N. in vitro diagnostic test for Placental Growth Factor

B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N. in vitro diagnostic test for Placental Growth Factor was recalled because customers have observed quality control values out of range at the lowest quality control level (qc level 1) for the pigf assay which require troubleshooting steps to resolve. if quality controls are not performed according to the instructions for use, the underestimation of qc1 may go undetected and could contribute to falsely elevated sflt-1/plgf ratio results. a falsely elevated sflt-1/plgf ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.. If you have this product, stop using it until you verify your lot or model details.

By RecallRadar Editorial Team

Sourced from FDA. Last verified April 20, 2026.

On this page

Free checker

Check if you qualify

See whether your device may qualify for a free repair or replacement program before you spend money.

Check if you qualify

Key facts

Date announced
April 15, 2026
Source agency
FDA

This recall involves safety concerns. Take action promptly.

Plain-English Summary

B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N. in vitro diagnostic test for Placental Growth Factor was recalled because customers have observed quality control values out of range at the lowest quality control level (qc level 1) for the pigf assay which require troubleshooting steps to resolve. if quality controls are not performed according to the instructions for use, the underestimation of qc1 may go undetected and could contribute to falsely elevated sflt-1/plgf ratio results. a falsely elevated sflt-1/plgf ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.. If you have this product, stop using it until you verify your lot or model details.

Is This Recall Still Active?

Status: Yes - Active

This recall is currently active. If your device matches the affected models, follow the official instructions as soon as possible.

What's Being Recalled

Quoted source excerpt from the official notice.

Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve. If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.

Quick Info

Key facts from the official notice, kept in the article flow for easy cross-checking.

Brand
Brahms gmbh
Category
medical
Source agency
FDA
Announced
April 15, 2026
Country
US
Status
Active

Who's Affected

Specific models and identifiers listed in the source notice:

Model #1

UDI: 04260157632196. Lot(Expiration): 59172(2026-11-23), 859075N(2026-11-23)

Technical model codes are manufacturer identifiers. Match both the product name and model code from your device settings, label, or packaging.

What You Should Do

Follow the remedy path from the official notice using the exact model and serial details above.

  1. 1Stop using the affected product until you verify whether your item is impacted.
  2. 2Check the official recall notice and match your model, lot, or serial details.
  3. 3Follow the official instructions for repair, replacement, refund, or disposal.
Open Official Recall Notice

Source & Verification

This page summarizes the official notice and keeps key facts easy to verify.

Official source
FDA
Announced by source
April 15, 2026
Last reviewed by RecallRadar
April 20, 2026

Frequently Asked Questions

Am I affected by this recall?

Check the affected models section on this page. The official notice currently lists UDI: 04260157632196. Lot(Expiration): 59172(2026-11-23), 859075N(2026-11-23).

What is the remedy for this recall?

Follow the official instructions for repair, replacement, refund, or disposal on the brahms gmbh recall page.

Where can I verify recall details?

Use the official source link in this page to confirm dates, eligibility, and next steps: https://api.fda.gov/device/enforcement.json?search=recall_number:Z-1795-2026&limit=1

Keep browsing with these internal hubs: brahms gmbh recall hub and Medical recall hub.

Official Resources

Use these primary sources for final eligibility and claim instructions.

View official notice from FDA

See an inaccuracy?

Help us improve the accuracy of this recall information.

500 characters remaining